Pre-installation requirements
51
2062973-003
•
7.2.3. Selecting and configuring S/5 Network
to see the configuration information you need to have available to configure the
patient monitor to the S/5 Wireless Network.
5.3.5 IX Network
•
The IX Network infrastructure shall be installed according to the CARESCAPE Network
Configuration Guide.
•
The installation site of the patient monitor shall have a wall jack and a network patch
cable for the IX Network.
•
Refer to the section
7.2.2. Selecting and configuring CARESCAPE Network
to see the
configuration information you need to have available to configure the patient monitor to
the IX Network.
•
Refer to the following sections for the information you need to have available for:
−
for IX printer configuration
−
−
−
7.16.1. Configuring the remote service
for Insite with Exc configuration.
5.4 Installing the mounting hardware
Ensure that all the applicable/ required mounting hardware is properly installed prior to the
installation of the patient monitor:
•
Mounting hardware for the patient monitor, either for a stand-alone installation or for an
installation to an anesthesia machine or to a ventilator.
•
Mounting hardware for the PSM module
•
Mounting hardware for the PDM module
•
Mounting hardware for the displays
•
Mounting hardware for the Unity Network ID connectivity device
NOTE: Refer to the patient monitor’s supplemental information manual for compatible
mounting hardware for each system component above.
WARNING
Use only manufacturer specified mounts.
5.5 Unity Network ID connectivity device installation
The Unity Network ID connectivity device shall be properly installed, configured and tested
according to the Unity Network ID Connectivity Device Service Manual prior to connecting it to
the patient monitor.
Make sure that the Unity Network ID connectivity device is configured as follows:
•
IP address is 192.168.253.x, where x is a number between 2 and 254.
•
Netmask is 255.255.255.0
•
The location of the Unity Network ID is set to a value other than the default (XXXX-XXX).
For example, BAY3|UNID3+.
Refer to the Unity Network Interface Device (ID) Service Manual for instructions on checking
and changing the IP address.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......