Configuration
95
2062973-003
7.19.2 Activating the installed software
After transferring the new software to the patient monitors, take the transferred, inactive
software into use by activating it.
Before you start:
•
Connect the patient monitor to the AC mains during software activation. Software
activation does not start if the patient monitor uses its battery.
•
Verify the compatibility of the connected bedside and network devices with the new
software version. Refer to the latest version of the CARESCAPE Modular Monitors
Supplemental Information Manual for a list of compatible network and bedside devices.
•
Contact GE Healthcare to get the latest version of the user and service documentation
•
Contact GE Healthcare for any inquiries regarding the software CD and/or activation
code for the new host software version.
NOTE: Loss of monitoring - Software is activated only when the patient monitor is in a patient
discharged / case reset state. Normal patient monitoring is unavailable until the software
activation is completed. This may take up to 10 minutes.
NOTE: The existing clinical and platform settings of the patient monitor are saved and are not
affected by the activation of the new patient monitor software version. However, any new or
changed clinical and platform settings in the activated patient monitor software version have
their factory default values and may require manual configuration. For more information, refer
to the latest version of the CARESCAPE Modular Monitors Supplemental Information Manual.
Contact GE Healthcare for any inquiries regarding the software CD and/or activation code for
the new host software version.
1.
Log in to Webmin.
2.
Select
Configuration
>
Software Management
.
The
Software Management
window displays.
3.
In the
Software List
, select the software that you want to activate::
4.
Select
Next
.
The selected
Software activation
window appears.
5.
According to the software you are activating, proceed to:
•
Activating host software immediately
•
Activating host software after next case end / discharge
•
Software
Purpose of use
Host Software
Patient monitor software for the host unit including:
•
PUIC SW for the DC/DC board
•
EMBC SW for the CPU Carrier board
PDM Software
CARESCAPE Patient Data Module
PiCCO Software
E-PiCCO module.
sGAS Software
CARESCAPE respiratory modules, E-sCAiOV family.
Summary of Contents for CARESCAPE Monitor B450
Page 12: ...CARESCAPE Monitor B450 4 2062973 003 For your notes ...
Page 16: ...CARESCAPE Monitor B450 8 2062973 003 For your notes ...
Page 46: ...CARESCAPE Monitor B450 38 2062973 003 For your notes ...
Page 56: ...CARESCAPE Monitor B450 48 2062973 003 For your notes ...
Page 62: ...CARESCAPE Monitor B450 54 2062973 003 For your notes ...
Page 108: ...CARESCAPE Monitor B450 100 2062973 003 For your notes ...
Page 124: ...CARESCAPE Monitor B450 116 2062973 003 For your notes ...
Page 136: ...CARESCAPE Monitor B450 128 2062973 003 For your notes ...
Page 175: ...Troubleshooting 167 2062973 003 ...
Page 180: ...CARESCAPE Monitor B450 172 2062973 003 For your notes ...
Page 218: ...CARESCAPE Monitor B450 210 2062973 003 For your notes ...
Page 224: ...CARESCAPE Monitor B450 A 4 4 2062973 003 Notes Used service parts Signature Date ...
Page 227: ...Maintenance check form B 3 4 2062973 003 Notes Used service parts Signature Date ...
Page 228: ...CARESCAPE Monitor B450 B 4 4 2062973 003 For your notes ...
Page 235: ......