163
Kinetics and affinity analysis 9
The subsets will be evaluated together, with rate or affinity constants that are
global for the whole data set and R
max
values that are global within each subset
but can differ between subsets.
Note:
The
Parameters
tab in evaluation of kinetics with a global R
max
parameter within subsets lists a parameter
i
n
which is set to 1 for the n
th
subset and 0 for all others. This is used to control R
max
values for each
subset: values are fitted as Rmax
×
i
n
, which returns a non-zero value for
the n
th
subset only. For affinity evaluations, each subset is represented by
one curve and R
max
is simply evaluated as a local parameter.
9.3
Batch mode evaluation
Batch mode allows you to evaluate multiple sample series (e.g. different
analytes interacting with a common ligand, or measurement of the same
interaction at different temperatures) automatically. You select the series to be
evaluated and the fitting model to be used, and evaluation is carried to
completion without further intervention.
To perform evaluation in batch mode, start the kinetics/affinity evaluation
(Section 9.2.1) and check the
Batch mode
option in the first dialog.
CAUTION!
Make sure that the same curve numbers are not assigned to
more than one data subset. If a curve is assigned to two data sets, the
software will try to evaluate the same curve with two different Rmax values,
and the fitting may be distorted.
Do not evaluate multiple copies of the same subset for multiple ligand
densities. Subsets that are duplicated will be weighted more than those that
are not duplicated in the fitting procedure.
Do not attempt to use subsets for any purpose other than multiple ligand
densities. The data will be evaluated in terms of multiple ligand densities
regardless of how you have assigned curves to the subsets.
Summary of Contents for Biacore T100
Page 1: ...GE Healthcare Biacore T100 Software Handbook ...
Page 2: ......
Page 8: ...8 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 13: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 13 Control Software ...
Page 14: ...14 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 30: ...3 Manual run 3 4 Ending a manual run 30 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 103: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 103 Evaluation Software ...
Page 104: ...104 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 207: ...Biacore T100 Software Handbook BR 1006 48 Edition AE 207 Appendices ...
Page 208: ...208 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 236: ...236 Biacore T100 Software Handbook BR 1006 48 Edition AE ...
Page 237: ......