Airway gas
17-13
Calibrating
Perform calibration every six months in normal use, once every two months in continuous use,
and whenever there are indications of errors in the gas readings to ensure that the
measurement accuracy remains within specifications.
Calibration setup
Required tools
•
P/N: 755534-HEL Calibration Gas Regulator
•
P/N: M1006864, Calibration Gas Regulator, US only
•
3 m / 10 ft anesthesia gas sampling line
•
For E-sCO, E-sCAiO, N-CAiO module:
−
P/N: 755583-HEL Calibration gas, CO2, O2, N2O, DESF, package of 1 can (with E-
sCAiO or N-CAiO module)
−
P/N: 755581-HEL QUICK CAL calibration gas, CO2, O2, N2O, package of 4 cans (with
E-sCO module)
−
P/N: 755571-HEL, Calibration Gas, 5% CO2, 54.5% O2, 36.0% N2O, 2.0%
DESFLURANE, BAL N2 (with E-sCAiO or N-CAiO module) US only
−
P/N: 755587, Calibration Gas, CO2, O2, Balance, 4 cans/pkg (with E-sCO module) US
only
NOTE: Use only the specified GE Healthcare calibration gas for the gas calibration to
ensure measurement accuracy. Do not use any other calibration gases. Check the
calibration gas container's labelling to ensure that the calibration gas has not expired.
NOTE: Ensure that the gas regulator is functioning properly before gas calibration.
Refer to the gas regulator's "Instructions for Use" letter for the annual maintenance
instructions.
Connections
1.
Ensure that the module is connected to the monitor.
2.
Ensure that you have a new D-fend or D-fend Pro water trap in use.
3.
Connect the gas regulator to the calibration gas container.
Figure 17-5 Attaching regulator to calibration can
4.
Connect a new gas sampling line to the sampling line connector in the water trap.
5.
Connect the other end of the gas sampling line to the regulator on the gas container.
Leave the regulator overflow port open to room air.
Summary of Contents for B20
Page 2: ......
Page 16: ...B40 B20 Patient Monitor xii ...
Page 17: ...1 Introduction ...
Page 18: ......
Page 27: ...2 System description ...
Page 28: ......
Page 58: ...B40 B20 Patient Monitor 2 30 ...
Page 59: ...3 Installation ...
Page 60: ......
Page 69: ...4 Monitoring basic ...
Page 70: ......
Page 79: ...5 Alarms ...
Page 80: ......
Page 102: ...B40 B20 Patient Monitor 5 22 ...
Page 103: ...6 Monitor setup ...
Page 104: ......
Page 123: ...7 Trends ...
Page 124: ......
Page 140: ...B40 B20 Patient Monitor 7 16 ...
Page 141: ...8 Print and record ...
Page 142: ......
Page 153: ...9 Cleaning and care ...
Page 154: ......
Page 163: ...10 Troubleshooting ...
Page 164: ......
Page 181: ...11 ECG ...
Page 182: ......
Page 205: ...12 Impedance respiration ...
Page 206: ......
Page 215: ...13 Pulse oximetry ...
Page 216: ......
Page 232: ...B40 B20 Patient Monitor 13 16 ...
Page 233: ...14 Non invasive blood pressure ...
Page 234: ......
Page 248: ...B40 B20 Patient Monitor 14 14 ...
Page 249: ...15 Invasive blood pressure ...
Page 250: ......
Page 261: ...16 Temperature ...
Page 262: ......
Page 267: ...17 Airway gas ...
Page 268: ......
Page 293: ...18 Entropy ...
Page 294: ......
Page 316: ...Abbreviations A 12 ...
Page 318: ...B40 B20 Patient Monitor B 2 ...
Page 319: ......