background image

 [EN]  English - 7

K386 Issue 1

Figure 3. (Part of figure) Procedure to transmit the Snapshot data

Snapshot - Contents of the Snapshot data

Figure 6: Example Snapshot data

Maintenance

This section gives procedures to maintain the unit in a good 
condition. Return the instrument to the supplier for all repairs. 

Maintenance – Clean the unit

Clean the case with a moist, lint-free cloth and a weak detergent. 
Do not use solvents or abrasive materials.

Maintenance – Replace the batteries  

 

To replace the batteries, refer to B1. Then re-attach the cover.
Make sure that the time and date are correct. The calibration 
facility uses the date to give service and calibration messages.
All the other configuration options stay in memory.

Maintenance – Recalibrate the unit

Note: GE Druck can provide a calibration service that is 
traceable to international standards.

Before you start the calibration, make sure that the time and date 
are correct (Table 1).
To get accurate results, do the calibration in stable temperature 
conditions, and to the applicable calibration standards.
Selection sequence: 

 Select Task (Table 2) 

 Set Up (Table 1) 

 Calibration 

.

When you select a channel, the display shows the applicable 
instructions to complete the calibration.

Table 5: .Calibration options

When the calibration is complete, select Calibration Due and set 
the new calibration date.

Calibration: Measure mV

The specified accuracy (refer to “Specification”) includes an 
allowance for temperature changes, reading stability for one 
year, and the uncertainty of the standard used for calibration. 
GE Druck recommend that you return the instrument to the 
manufacturer or an approved service agent for calibration.
If you use an alternative calibration facility, make sure that it 
uses these standards:

Accuracy of the mV calibration equipment:
0.008% r 0.003% FS or better.

Calibration limit for the instrument (mV measure): 0.01% 
r 4 µV or better.

Step

Procedure

A1 - Item [9a] (Optional): Connect the 5-pin connector 
to the IDOS instrument.

A1 - Item [9a] (Optional): Connect the 9-pin connector 
to the serial port on the PC or to a serial printer (refer 
to the PC or printer instructions).

PC: Make sure that Windows

®

 HyperTerminal or 

similar program is set up to use the specified data 
format (refer to your PC on-line Help).

Serial printer: Make sure that the printer can use the 
specified data format.

 

Data format

Baud rate: 19200
Parity: None 

Data bits: 8 
Stop bits: 1

Software handshake: Xon/Xoff Flow Control
ASCII format text; Comma delimited

Transmit the data to the PC or printer (Table 4).

PC: Use Windows

®

 HyperTerminal or a similar 

program to make a text file. You can then import the 
text file into a spreadsheet.

Model

820

Serial No

8200000001

Calibration due

01/01/2004

Tag

H2O-1

Task Id

110

T1 in

Date

Time

°C K T1

08/07/2004

12:06:44

27.6

08/07/2004

12:06:46

27.2

Tag

B145

Task Id

164

T1 In

Tare On

T2 In

Filter On

Tare On

Date

Time

°C K T1

°C K T2

08/07/2004

14:41:36

32.5

32.4

08/07/2004

14:41:38

32.5

32.4

08/07/2004

14:41:41

32.4

32.5

1

Display: Enter 

Calibration PIN

2

3

Menu:

Select Channel

4

5

+

 

Factory PIN = 4321

(Table 5)

Options

Description

mV... 

To calibrate the specified thermocouple channel.

CJ...

To calibrate the specified cold junction channel. 

 IDOS ...

UMM - Pressure only. To calibrate the specified IDOS 
channel. Refer to the UMM specification

To set the date for the next calibration. 
After the specified calibration date, there is a warning 
message. There is a selection box to stop the warning.

To change the calibration PIN (Personal Identification 
Number).

B1

Summary of Contents for DPI 820

Page 1: ...K386 Issue 1 DPI 820 Dual Input Thermometer User manual 99 Washington Street Melrose MA 02176 Fax 781 665 0780 TestEquipmentDepot com ...

Page 2: ...6 Issue 1 A1 A2 10 1 3 4 2 5 6 7 9 8 A 2 11 B1 12 10 10a 9a 17 18 19 20 14 15 21 16 22 13 B1 Test Equipment Depot 800 517 8431 99 Washington Street Melrose MA 02176 FAX 781 665 0780 TestEquipmentDepot com ...

Page 3: ...ures the instructions for the UMM if applicable and this publication Before you start an operation or procedure make sure that you have the necessary skills if necessary with qualifications from an approved training establishment Follow good engineering practice at all times WARNING It is dangerous to ignore the specified limits for the instrument or to use the instrument when it is not in its nor...

Page 4: ...S232 cable 9a Optional item Part No IO800D RS232 cable 5 pin to 9 pin D type connector Use to copy the snapshot records to a PC or to a printer Refer to Operation 10 Thermocouple connector Refer to Operation 10a Type K thermocouple two included 11 Connection point for some of the optional accessories Refer to the datasheet 12 Battery compartment Refer to B1 A1 Item Description 13 UMM Pressure only...

Page 5: ...e To select the applicable international temperature scale IPTS 68 or ITS 90 To select and set up the backlight facility timer Additional data Select Settings To select and set up the power off facility timer Additional data Select Settings To show the battery level To set the display contrast Increases decreases To set the time date The calibration facility uses the date to give service and calib...

Page 6: ... Settings 4 5 6 Settings x y Table 3 Sk1 Edit Options Description To include maximum minimum and average values for the measurement task If applicable Only available when there is one input measurement task To select and set up a tare value for the measurement task a specified value or the reading on the display The instrument subtracts a positive tare value and adds a negative tare value Addition...

Page 7: ... temperature with one or two thermocouples The examples in Figure 1 show how to compare temperatures or measure different temperatures in a system Figure 2 Example configuration To measure the temperature difference Operation IDOS Pressure measurements UMM Pressure options only Read all the instructions supplied with the UMM and then use the specified procedures to connect it Figure 3 4 Figure 3 E...

Page 8: ...s to see the values that are set up If necessary adjust the values Table 4 Snapshot functions Snapshot Use the Snapshot data Use HOLD to record the data Table 4 When you have sufficient records select Stop Log You can then read the data back on the display or transmit it to a PC or printer Figure 5 Part of figure Procedure to transmit the Snapshot data Settings K T1 Menu Set Up 2 3 Menu Snapshot 4...

Page 9: ...ibration If you use an alternative calibration facility make sure that it uses these standards Accuracy of the mV calibration equipment 0 008 reading 0 003 FS or better Calibration limit for the instrument mV measure 0 01 reading 4 µV or better Step Procedure A1 Item 9a Optional Connect the 5 pin connector to the IDOS instrument A1 Item 9a Optional Connect the 9 pin connector to the serial port on...

Page 10: ...4 180 70 0 9 0 5 T IEC 584 94 752 70 400 0 6 0 3 B IEC 584 122 932 50 500 7 2 4 0 B IEC 584 932 2192 500 1200 3 6 2 0 B IEC 584 2192 3308 1200 1820 1 8 1 0 R IEC 584 58 32 50 0 5 4 3 0 R IEC 584 32 572 0 300 3 6 2 0 R IEC 584 572 3216 300 1769 1 8 1 0 S IEC 584 58 32 50 0 4 5 2 5 S IEC 584 32 212 0 100 3 4 1 9 S IEC 584 212 3216 100 1769 2 5 1 4 E IEC 584 454 238 270 150 1 6 0 9 E IEC 584 238 1832...

Reviews: