- 7 -
Figure 5 Test menu interface
In the "test current" position click mouse to switch the current value between 50A, 100A;
rotate the mouse to the "test time" position, using a rotary mouse input data, you can set the test
time. Note: The test time setting range: 5s ~ 599s, when beyond the scope, the system returns to the
default value: 10s. In order to ensure more accurate test results, recommended test time using
default values 10s.
Click on the "Test" key, the system enters the "Test Results" screen.
Click on the "return" key, the system returns to the previous screen.
7. Test Results
Click on the "Test Menu", then "test" to enter into the "Test Results" screen, as shown in
Figure 6. Sequentially displayed is resistance value, test current value and test time on screen. Note:
At this point the current line, there are large current flows, the current line must not be forcibly
pulled out, otherwise it may cause harm to the operator and the instrument.
Figure 6 Test Results - Test
When testing, the system displays the "stop", "Return" key. Click on "Stop", the system stops
counting, the current stops output. Click on "Back", the system stops counting, the current stops
output and return to the previous screen. Note: The first few seconds of the beginning of the test,