ZXFC MOV SPD TESTER

- 5 -

After clicking the "test" button, the test voltage starts at the rate of 100V/s and step up from the voltage preset

value. After the green light is on:

If the buzzer does not alarm, the display is the ignition voltage of measured discharge tube in the

measurement range.

If the buzzer alarms, the display is the upper limit of the measurement range. At this time, the ignition voltage

Vsdc value of the tested product is greater than the upper limit of the measurement range (upper limit range)

and no ignition. Increase the "upper limit of test range" and repeat the test again.

The green indicator lighting up time along with the value display lasts for about 2 seconds, and then revert to

the preset state.

When the green indicator light is on, remove the measured discharge tube and wait the preset voltage is

restored then make the next test.

Continuous measurement

Turn single / continuous to low position to start continuous measurements

Self-check and others

A) Varistor/MOV 1mA value test

Select varistor / discharge tube to varistor, open high voltage, preset voltage to above 10V, short circuit "+"

“-”ends, display should be "000", long press "display conversion" key, display should be "1000", otherwise it

indicates the instrument has problems, please contact the after sale personnel in time.

B) examination of the 0.75U1mA value of the varistor test

Make the test ends open circuited and press test key, the display screen should show the upper limit of the

range. When the green light is on, press "display conversion" key, the display should be 0.75 times of the upper

limit of the range.

C) in the process of testing, long press "display conversion” to display the value under 1mA (1000uA) test

condition during U1mA measurement, or the value of 0.75U1mA test condition during I0.75U1mA

measurement.

Others

Use voltage preset and range adjustment function the instrument can be used as 1999V/1mA DC stablized

power supply and 1mA constant current source (the instrument constant output 1mA test current when the

load current reaches 1mA). With the use of "display conversion" key, the value of load voltage and current (V/I

characteristics) can be measured.

3. completion of the test

Summary of Contents for ZXFC

Page 1: ...ZXFC Lightening Protection Devices Tester ...

Page 2: ...ZXFC MOV SPD TESTER 1 MANUAL I Product Functions and Features 2 II Technique Parameters 2 III Panel 3 IV Operation Method 4 V Precautions 6 VI After sale service 6 VII Packing list 7 ...

Page 3: ...good reliability 5 Audible alarm for over range during the measurement after presetting 6 Continuous measurement can be used for batch testings 7 Simple panel easy operation light weight easy carrying II Technique Parameters 1 MOV measurement Technique parameter range error condition Initial action voltage U1mA 0 1999V 1 1d 1mA 5μA Leakage current 0 75U1mA 0 199 9μA 1μA 1d 0 75U1mA 1 1d 2 SPD meas...

Page 4: ...t Rear 1 HV limit 2 Single Continuous 3 MOV SPD 4 HV indicating light 5 HV on off 6 Display conversion 7 Test 8 Testing light 9 Voltage preset 10 Screen 11 Buzzer 12 power switch 13 Test terminal red 14 Test terminal black 15 Power socket ...

Page 5: ...gle continuous to lower side connect to the testing sample after made the preparation mentioned above Press HV on off key and test key the test voltage rises from the voltage preset value at the rate of 100V s when the green indicator lights up the screen displays the voltage which is the ignition voltage of the tube 2 screening method Turn varistor discharge tube to low position discharge tube si...

Page 6: ...or MOV 1mA value test Select varistor discharge tube to varistor open high voltage preset voltage to above 10V short circuit ends display should be 000 long press display conversion key display should be 1000 otherwise it indicates the instrument has problems please contact the after sale personnel in time B examination of the 0 75U1mA value of the varistor test Make the test ends open circuited a...

Page 7: ...han the upper limit voltage Otherwise the test voltage will be limited in the range upper limit in the discharge tube test the test indicator will be repeatedly lit and accompanied by over range sound prompts At this time lower the preset voltage 4 The test voltage of this machine is as high as 1900V keep the panel and work table clean and dry to avoid the error caused by leakage of current arc an...

Page 8: ...MOV SPD TESTER 7 VII Packing list 1 Main tester 1 pcs 2 Test lines 1 set 3 test clips 2 pcs 4 special power supply 1 pcs 5 Manual 1 copy 6 Inspection reports 1 copy 7 Warranty card 1 pcs 8 Carrying box 1 pcs ...

Reviews: