GDZX ZXFC Manual Download Page 1

ZXFC Lightening 

Protection Devices Tester

Summary of Contents for ZXFC

Page 1: ...ZXFC Lightening Protection Devices Tester ...

Page 2: ...ZXFC MOV SPD TESTER 1 MANUAL I Product Functions and Features 2 II Technique Parameters 2 III Panel 3 IV Operation Method 4 V Precautions 6 VI After sale service 6 VII Packing list 7 ...

Page 3: ...good reliability 5 Audible alarm for over range during the measurement after presetting 6 Continuous measurement can be used for batch testings 7 Simple panel easy operation light weight easy carrying II Technique Parameters 1 MOV measurement Technique parameter range error condition Initial action voltage U1mA 0 1999V 1 1d 1mA 5μA Leakage current 0 75U1mA 0 199 9μA 1μA 1d 0 75U1mA 1 1d 2 SPD meas...

Page 4: ...t Rear 1 HV limit 2 Single Continuous 3 MOV SPD 4 HV indicating light 5 HV on off 6 Display conversion 7 Test 8 Testing light 9 Voltage preset 10 Screen 11 Buzzer 12 power switch 13 Test terminal red 14 Test terminal black 15 Power socket ...

Page 5: ...gle continuous to lower side connect to the testing sample after made the preparation mentioned above Press HV on off key and test key the test voltage rises from the voltage preset value at the rate of 100V s when the green indicator lights up the screen displays the voltage which is the ignition voltage of the tube 2 screening method Turn varistor discharge tube to low position discharge tube si...

Page 6: ...or MOV 1mA value test Select varistor discharge tube to varistor open high voltage preset voltage to above 10V short circuit ends display should be 000 long press display conversion key display should be 1000 otherwise it indicates the instrument has problems please contact the after sale personnel in time B examination of the 0 75U1mA value of the varistor test Make the test ends open circuited a...

Page 7: ...han the upper limit voltage Otherwise the test voltage will be limited in the range upper limit in the discharge tube test the test indicator will be repeatedly lit and accompanied by over range sound prompts At this time lower the preset voltage 4 The test voltage of this machine is as high as 1900V keep the panel and work table clean and dry to avoid the error caused by leakage of current arc an...

Page 8: ...MOV SPD TESTER 7 VII Packing list 1 Main tester 1 pcs 2 Test lines 1 set 3 test clips 2 pcs 4 special power supply 1 pcs 5 Manual 1 copy 6 Inspection reports 1 copy 7 Warranty card 1 pcs 8 Carrying box 1 pcs ...