background image

- 2 -

Safety Precautions

If you are using this product for the first time, please pay attention to the

following:

At the beginning of measurement, first open the needle valve on the

instrument measuring pipe, and then adjust the flow with the flow valve on the

panel. If the measurement is finished, the operation will be reversed.

When the instrument is not used for a long time, there will be some air left

in the test pipe and air chamber. Therefore, during the first test, the residual air

will affect the test speed, because the impurities in the air must be removed

before the tested SF6 gas reaches the balance. So we will find that when we

test the second SF6 Electrical equipment and measure it later, the speed will

be very fast (1-2 minutes).

In case of high temperature, we suggest to arrange the measurement

time as far as possible when the temperature is low in the morning, because

the higher temperature will affect the accuracy of measurement.

In the working environment, the product may cause radio interference. In

this case, it may be necessary for users to take practical measures for their

interference.

Please read and keep this manual for future reference.

Check whether the equipment is abnormal before opening.

Summary of Contents for ZXCD SF6

Page 1: ...ZXCD SF6 Purity Analyzer ...

Page 2: ...cting SF6 devices 4 2 Startup initialization 4 3 Check power 5 4 Start measuring 5 5 Store data 6 6 View data 6 7 Delete data 7 8 Time Setting 7 9 Calibration data 8 10 After measurement 8 Ⅳ Matters needing attention 9 Common faults and troubleshooting 9 Ⅵ Common faults and troubleshooting 11 ...

Page 3: ...eed because the impurities in the air must be removed before the tested SF6 gas reaches the balance So we will find that when we test the second SF6 Electrical equipment and measure it later the speed will be very fast 1 2 minutes In case of high temperature we suggest to arrange the measurement time as far as possible when the temperature is low in the morning because the higher temperature will ...

Page 4: ...fe and reliable without air leakage 5 Data storage it adopts large capacity design and can store up to 1000 groups of test data 6 Clear display the 5 inch large touch LCD directly displays the current purity gas flow ambient temperature ambient humidity time and date etc 7 The built in USB interface can be connected with the PC and the internal data of the instrument can be uploaded to the compute...

Page 5: ...more than 10 hours working temperature 30 70 size 280 150 300 mm weight 5 kg Ⅲ Operation Method 1 Connecting SF6 devices Connect the screw end of the measuring pipe with the switch joint and tighten it with a wrench Close the flow regulating valve on the front panel of the main engine Insert one end of the quick connector on the test pipe into the air inlet on the instrument Connect the exhaust pi...

Page 6: ...nt can be used while charging but the charging time will be greatly extended 4 Start measuring 1 When the power is turned on the instrument will initialize automatically After self calibration it will enter the measurement state automatically 2 Confirm that the high precision control valve on the instrument is closed and test the quick plug of the pipeline 3 Connect the other end of the test pipe ...

Page 7: ... menu text on the screen call up the main menu interface select print data you can print out the current test data select save data you can enter the data saving interface Press the red configuration button to input the test number After confirmation press save data in the lower right corner to save the current data to the instrument The instrument can save up to 1000 pieces of data 6 View data ...

Page 8: ...utton to print the currently viewed history 7 Delete data Select delete data function in the main menu interface to enter the delete data interface At this time press the trash can icon to confirm the deletion of all data Press the arrow in the lower left corner to undo the deletion and return to the main interface Note data cannot be recovered after confirming deletion 8 Time Setting ...

Page 9: ...ecause of changing the calibration data at will it will seriously affect the use of the instrument So we do not open the data calibration function If necessary please contact our technical support 10 After measurement 1 Close the high precision needle type regulating valve on the instrument 2 Remove the adapter from the SF6 Electrical equipment 3 Remove the quick plug between the test tube and the...

Page 10: ...e it is recommended that the instrument be able to power every 1 2 months when it is not used for a long time To keep the lithium battery active 5 The instrument shall not be put into the aluminum alloy packing box and shall be placed on the test bench or instrument rack to prevent dust and moisture 6 The instrument shall be calibrated with standard gas once a year It can be sent to the manufactur...

Page 11: ...for maintenance Charging indicator is not on Circuit fault Check circuit Battery voltage too low Charge Buzzer does not sound Buzzer failure Replace buzzer Circuit fault Check the buzzer circuit The buzzer keeps ringing Circuit fault Inspection instrument Sensor output line loose Check the sensor cable LCD no display Poor LCD connection Check the connector Voltage below 5 8V Charge or replace the ...

Page 12: ...and troubleshooting NO Name Qty Remarks 1 Host 1set 2 Inflatable pipe with needle valve 1pc 3 Outgassing pipe 1pc 4 charger 1pc 5 Transition joint option 1pc 8 7 manual 1pc 8 Test report 1pc 9 Certificate warranty card 1pc ...