background image

11. Press T to adjust the 20mA output by measuring the mV across TP8 and TP9. Press H or L to adjust and then press 

SPACE to set and exit. 

12.  Before using span gas, ensure that an autozero is not about to take place. This is indicated by the MPU LED 

flashing at 1 Hz. The MPU LED on the PCB will flash normally at 1 Hz but when there is less than 30 seconds to an 
autozero flash rate will increase to 2Hz i.e faster flash.

13.  Apply a test gas to the sensor (typical - 100ppm Isobutylene) for 1 minute at a flow rate of 1 litre/min. Use the 

UP and DOWN buttons on the PCB to give correct  4-20mA level. (ie. 5.6mA for 100ppm). Remove gas and wait 
5 minutes. If a maximum range gas is available, apply for 1 minute and note the value of A. This can then be 
entered via Hyperterminal using the M command giving better accuracy across the range. Remove gas and wait 2 
minutes. 

14.  Connect the sensor to a Combi alarm panel and ensure that it reports in correctly.  

Note: 

Fit the end of line (EOL) link J1 if  the sensor is to be installed at the end of the sensor cable.

15.  If a display board is fitted via connector J5 and U12 adjust VR1 for LCD contrast.

16.  Remove J9 test link for normal operation. 

17.  Ensure J29 address link is removed (this is only used when changing address from a panel)

Manual calibration (without PC or LCD)

18. Zero - with no gas applied, measure between Ag and Sig (Y) and adjust VR2 to give 50mv. 

With J9 test link inserted, check the 4-20mA output to be 4mA as measured at TP8 to TP9 = 4mV

19.  Span - apply test gas (typical 50% of range) and use UP and DOWN buttons on the PCB to give correct 4-20mA 

output. (50% gas = 12mA).

20. Remove gas and ensure the output returns to the 4mA level and then remove the J9 test link

 

Setup using magnets - LCD

WIN versions of the sensor have an LCD display which shows the Gas and range together with the sensor address and 
a display of the number of seconds until an autozero will occur.  
Also incorporated are 3 reed switches which can be activated using external magnets through the glass window of the 
flameproof XDIwin enclosure. These magnets do not act instantly and have to be in close proximity to L, M and R on the 
front display for a few seconds to activate a software setup function.

The right magnet allows the CAN address of the sensor to be changed. When the address menu is displayed with a 
prompt to remove the magnet, and then the display shows the address and that the right magnet will decreases it whilst 
the left magnet will increase it. This is then stored in internal non-volatile memory and the display will automatically 
revert to normal operation. 

The centre magnet is used to inhibit the sensor. As with the left and right magnet functions the display requests that you 
remove the magnet and then the state of inhibit appears on the LCD.  
The left magnet then puts the sensor into inhibit whilst the right magnet removes it. The amber LED on the front panel 
under the LCD flashes when the sensor is inhibited. When all magnets are removed, the display will revert to normal 
operation. The direction of the alarms is displayed as ^ for rising and v for falling but these can be changed using left 
and right magnets together. 

The left and right magnets together allow the calibration menu to be used. Removing both magnets as instructed on 
the LCD presents the first part of this multi menu which is ZERO. With no gas present the display shows the PPM value 
but also typically [0.02] This number between the square brackets shows how close the autozero has achieved. [0.00] is 
definite zero and the PCB led D1 will be flickering orange. 

A 15 second timer is displayed and is reset back each time a magnet is near. Waiting till timeout is acceptable but this 
timeout can be speeded up by placing a magnet near to the centre position. 

SPAN is the next part of the menu and gas should be applied to the sensor at this time (recommended - 100ppm 
Isobutylene - see GDS application notes C645). A 2 minute counter is displayed before the next part of menu is 
activated. The left magnet now increases the PPM reading and the right magnet reduces. 

Summary of Contents for XDI-XDIwin/30J

Page 1: ... bypass 4 20mA terminal 5 Set the sensor voltage to 5 0V measured across SV and 0V adjust by slowly turning RV4 clockwise DO NOT EXCEED THE REQUIRED VOLTAGE OTHERWISE PERMANENT DAMAGE WILL OCCUR 6 Insert jumper G2 for a nominal gain value G1 is lowest gain G5 is highest or a combination G1 to G5 7 Connect a PC to J3 with a sensor programming module and a USB to RS232 converter part no 160510 and p...

Page 2: ...reed switches which can be activated using external magnets through the glass window of the flameproof XDIwin enclosure These magnets do not act instantly and have to be in close proximity to L M and R on the front display for a few seconds to activate a software setup function The right magnet allows the CAN address of the sensor to be changed When the address menu is displayed with a prompt to r...

Page 3: ...dress and serial number are then output to the PC together with alarms and calibration date etc This also starts 1 minute autozero which is useful when adjusting VR2 zero potentiometer Pressing C on the PC will list the available commands V Allows a view of the gas log taken at one minute intervals over 48 hours A To set the sensor address G Select gas type Select the gas type from a list Z Prompt...

Page 4: ...pecification may be modified at any time without prior notice Technical Sheet ref C1518B XDI XDIwin 30J PID SENSOR Set up procedure 292 D6C Issue Bv5 Figure 1 Figure 2 Figure 3 Appendix 2 The PID sensor can react to gas flow and humidity When gas is removed from the sensor the displayed PPM gas value may rise and take some time to return to normal Autozero is therefore delayed by 2 minutes after t...

Reviews: