GASMAX II Operation & Maintenance Manual, Revision 7.0
Page 15
LOCAL SMART TOXIC SENSOR WIRING (CH1)
Local toxic sensors are factory installed in a sensor head directly attached to the GASMAX II enclosure.
Local toxic sensors are always connected to the Channel 1 Smart Sensor connector
“ ”
located on the
back of the Display Assembly
(See Fig.4-5).
IMPORTANT
: TOXIC SENSORS MAY FAIL TO OPERATE PROPERLY IF POWER IS NOT APPLIED WITHIN 6
MONTHS OF SHIPMENT. CONTACT GDS CORP FOR RECOMMENDATIONS IF THIS SITUATION MAY OCCUR.
Figure 3-7: Local Smart Toxic Sensor Connection
ibbon to isplay
I Power upply
Bridge ensor
C
M
C
T
P
T
T
G
TB
TB
ibbon to isplay
isplay ssembly
ocal Toxic ensor
Cable from sensor head connects to
on back of display assembly
Do not connect signal or power
wiring to terminal block TB1.