GASMAX CX Operation & Maintenance Manual, Revision 1.0
Page 41
DIAGNOSTICS
MENU
The
Diagnostics
page
provides
tools
for
use
during
setup
or
testing.
Tests
for
optional
features
are
not
available
if
the
feature
is
not
installed.
Figure 10-7: Diagnostics Menu Tree
!! Warning !!
Diagnostic mode.
Alarms will not
Be processed.
Edit
(OK)
Next (Cancel)
Diagnostics
Relays
→
Analog Inputs
→
Analog Outputs
→
LED Test
→
Serial Ports
→
ADC Readings
→
Relays
Relay 1 Off
Relay 2 Off
Relay 3 Off
Fault Relay On
Ch.1 EC - Loc
Counts: 856
Temp: 24.5 degC
—————————————————
Ch.2 4-20mA
Counts: 0
Current: 0.00mA
Analog Outputs
4-20mA Out1 0mA
4-20mA Out2 0mA
Out1 Fdbk 0.07
Out2 Fdbk 0.06
LED Test
Comm Test
Connect Loopback
Port 1 => 2: Bad
Port 2 => 1: Bad
420 Out1: 4.60mA
Fdbk 1: 4.57mA
420 Out2: 4.05mA
Fdbk 2: 4.14mA
Sensor V: 0.01V
SensrAmp: 0.78V
PSU: 22.93V
WARNING
–
ENTERING
DIAGNOSTICS
MODE
DISABLED
ALARM
FUNCTION
AND
REQUIRES
A
DEVICE
REBOOT
ON
EXIT.
Relays
Menu
Allows
individual
relays
to
be
turned
on
or
off
under
user
control.
Only
available
if
optional
Relay
/
Modbus
board
is
installed.
Analog
Inputs
Menu
Shows
current
input
readings
for
channel
1
and
channel
2.
Analog
Outputs
Menu
Shows
the
live
output
values
on
analog
channel
1
and
analog
channel
2.
Also
shows
the
feedback
monitor
reading
for
both
channels.
LED
Test
Menu
Cycles
the
external
LED.
Comm
Test
Menu
Performs
a
loopback
test
to
verify
proper
operation
of
the
serial
components.
Only
available
if
optional
Relay
/
Modbus
board
is
installed
.
ADC
Readings
Displays
values
monitored
by
internal
analog
‐
to
‐
digital
converters.