GASMAX/TX Operation & Maintenance Manual, Revision 1.0
Page 37
CHANNEL CONFIGURATION MENU
The Channel Configuration menu contains settings that are unique to each channel. There are two
channels and each menu contains identical entries.
NOTE:
WHEN EDITING CHANNEL INFORMATION, MAKE SURE TO SELECT THE CORRECT CHANNEL MENU.
Figure 11-2: Channel Configuration Menu
CHANNEL ENABLED / DISABLED
–
Channels should only be enabled if a sensor is attached.
SENSOR TEMP
– Shows the current sensor temperature.
MEASUREMENT NAME
may be edited to contain any 16-character field and is typically used to describe
the monitored point by user tag number or other familiar terminology. To modify the existing setting,
press the EDIT key when the cursor is pointing to the entry. Use the UP or DN keys to change the
character, and the NEXT key to move to the next character. Press EDIT again when done.
EUNITS
(engineering units) may have up to a 10 character ASCII field. Many common gases have pre-
configured Eunits based upon the sensor type and each may be edited in this menu.
ZERO (0%)
defines the reading to be displayed when the input is at 0%.
SPAN (100%)
defines the reading to be displayed when the input is at 100%. The highest reading allowed
is 9999 including negative polarity sign. Polarity is only indicated for negative readings.
DECIMAL POINTS
set the resolution of the LCD readings and may be set for 0, 1 or 2. For example, ZERO
readings for 0, 1 & 2 decimal points displays as “0”, “0.0” & “0.00”.
CAL SPAN VALUE
sets the engineering units value of the calibration gas. GDS Corp recommends that the
calibration gas value be between 25% and 75% of full scale. Calibrating at 100% of scale is not
recommended.
READOUT DEADBAND
forces low values to continue to read zero. This is useful when there are small
amounts of background gases that may cause fluctuating readouts. The highest amount of deadband
allowed is 5%.
CHx CONFIG PG1/2
CHx CONFIG PG1/2
CHx CONFIG PG1/2
CHx CONFIG PG1/2
Measurement Name
Eunits ppmH2S
Zero (0%) 0
Span (100%) 100
Decimal Points 0
Cal Span Value 50
Readout Deadband 1%
Track Negative NO
Backup Config
Restore Config
Load Defaults
CHANNEL x
CHANNEL x
CHANNEL x
CHANNEL x
Channel ENABLED
Configure Readout
Alarm Settings
Sensor Information
Technicians Only
Sensor Temp 29.0C