and about 0.500V is displayed when full scale of gas
is detected on sensor. The value can vary due to
calibration (Span & Zero) by program when
sensitivity is changed due to aging of sensor)
-. Touch “RESET” switch to go back to “CALIBRA.
MODE”.
9. S/W version check
9.1 S/W version check
-. Touch "FUNC" switch with magnet bar for over 2
seconds and detector will go into PASSWORD MODE.
[PASSWORD]
-. Input password (2 digits set by user) with “
▲
” or
“
▼
” switch and touch “FUNC” switch to go into
Menu selection mode
-. Select “S/W VERSION MODE” by touching “
▲
” or
“
▼
” switch
[ V E R S I O N ]
[ G T D - 2 0 0 0 ]
-. When S/W version mode is displayed, touch
“FUNC” switch to go into S/W version mode
-. GTD-2000 is product model name and ‘V1.00’
represents S/W version.
-.Touch “RESET” switch to return to “S/W VERSION
MODE” Menu.
10. MAINTENANCE
Most of maintenance functions were set up at factory, so be careful not to operate
this function at normal condition. And in case of unavoidable situation, please
manipulate this function with technical assistance from GDA.