4/59
EN
4. PERSONNEL INSTRUCTIONS
The Medical Devices Directive 93/42/EEC states that product provider must ensure that all
personnel handling the product are provided with the operating instructions & performance
data.
Do not use the product without properly familiarization of the product and its safe operation
as defined in this Instruction for use. Ensure user is aware of particular information and knowl-
edge required for the gas in use.
5. PRODUCT DESCRIPTION
Combination valve combines the function of shut-off valve of a high pressure gas cylinder and
pressure regulator for the use with medical gases. Gas from the cylinder is first controlled by
the main shut-off valve and then passed through the pressure regulator and finally delivered to
the patient through the flow outlet or the pressure outlet. Outlet positive pressure is fixed by the
manufacturer and each combination valve is provided with a low-pressure relief valve to protect
against undesirable exceeding of outlet pressure.
There are two basic alternatives:
•
Valve outlet via calibrated nozzles, outlet flows can be changed by the control knob
•
Valve outlet via quick-coupler (the outlet flow is a constant one) in combination with valve
outlet via calibrated nozzles where outlet flows can be changed by the control knob
K
Serial number
Inlet connection
type
Reference
number
Inlet pressure
End of
lifetime
Filling port type
Compliance with
2010/35/EU
Compliance with
93/42/EEC
MediVitop® marking detail
20 kg
Total weight
allowance of the
package
THE LAYOUT OF THE TEXT IS ONLY ILLUSTRATIVE, SOME VARIANTS MAY BE A DIFFERENT LAYOUT