HiToP 1touch
27
gbo Medizintechnik
AG
Version 1.2
6.2
Disposal of the device and the accessories
According to the WEEE Directive 2002/96/EG (waste electrical and electronic
equipment) this device must not be disposed of with the domestic waste. The device
must be returned to the manufacturer for disposal. The manufacturer is committed to
guarantee the disposal of all devices marketed. This is also indicated by the WEEE
sign (crossed out waste container).
6.3
Cleaning and disinfection
6.3.1
Cleaning the Device
Clean the device with a damp cloth. Never use abrasive cleaners.
Fingerprints on the touchscreen should be cleaned with a dry cloth or some
isopropanol.
Warning!
Unplug the mains plug out of the socket prior to cleaning and
disinfection.
The device is not suited for heat sterilization or for the sterilization
with gases.
The
HiToP
1touch
is suited for wiping disinfection. Make sure that no liquids soak
into the device. Under no circumstances the plug or socket must get wet. Do not
sprinkle the device for cleaning or disinfection. The device is not suited for hot
sterilisation or sterilisation with gases.
6.3.2
Cleaning the Elastic Bands
The elastic straps can be washed in a washing machine. Please observe the washing
instructions of the instruction label or observe the washing instruction on the
packaging.
6.3.3
Cleaning the Electrodes
You can clean the rubber electrodes after a treatment with some warm water (appr.
40°C) and soap suds on the conductive (black) side. Dry the electrodes with a cleaning
cloth or let them air-dry. The cleaning process removes the film left by the Aloe spray or
the contact spray and the electrodes will regain a good conductivity.
!!
Note!
The conductivity of the electrodes is optimized by the use of a certain
amount of graphite. Consequently, when using and cleaning the
electrodes, black color may come off.