8 Maintenance
Functionality, reliability, and safety characteristics of the current stimulation device are only guaranteed in
case of proper use in accordance with the operating instructions. Safety control, maintenance work, repair
work, and modifications shall be carried out only by the manufacturer or by service agents authorized by
him. In case of a failure, parts which influence the safety of the device shall be only replaced by original
spare parts of the manufacturer. The electric installation must correspond to the requirements in accordance
with VDE/IEC.
The device does not contain any parts which need maintenance work done by the user.
8.1 Safety Controls
The device is subject to the provisions of the Medical Device Directive. The safety controls have to be
carried out on the basis of this directive. Thereby, the operator regulation has to be especially observed.
Irrespective of the legal rules or beyond the scope of the Medical Device Directive, it is recommended to
have the device checked by the manufacturer or by a service agency authorized by him at 12-months
intervals.
The check shall consist of at least the following criteria:
•
Electrical safety check in accordance with the test plan of the manufacturer,
•
Check of the device in respect of external integrity,
•
Check of all display and operating elements in respect of damages,
•
Check of all inscriptions in respect of legibility.
8.2 Cleaning, Disinfection and Care
For cleaning and disinfection of the current stimulation device and its accessories (except electrode fleece,
electrode pockets or felts, see chapter 8.2.1) there should not be used any agents containing higher portions
of phenol derivatives, alcohol, compounds of chlorine or peracetic acid. It is recommended to use
disinfectants on aldehyde basis.
The device is not suited for heat sterilization or for sterilization with gases.
Warning!
The current stimulation device is suited for wiping disinfection. It has to be observed that no liquids enter
the device. In no case shall the plug or socket get wet. For cleaning or disinfection the device may not be
drizzled.
8.2.1 Electrode Fleece, Electrode Pockets or Felts
1. Put the material into neutral soap-suds and rub it slightly. If necessary, disinfect with disinfectant
solution, but do
not
put it into it.
2. Leave the parts over the night in the neutral soap-suds or tap water.
3. Take them out, and wash them several times thoroughly out under cold or warm running water.
4. Let them drop, but do
not
wring them out.