DUODYNATOR
®
1049
11
gbo
Medizintechnik
AG 2008
Version 1.3
2 Start of Operation
2.1 Transport and Assembly
The current stimulation device is a portable unit. There is a carrying handle in the form of a recessed grip in
the base plate. To place the unit, each flat surface is appropriate. A wall distance of at least 20 cm has to be
provided. The device should not be placed in front of radiators.
The current stimulation device
corresponds to the regulations DIN/VDE 0750, EN 60601. It is a device of
protection class I. Within the scope of the Medical Device Directive (MDD) the current stimulation device
belongs to class IIa (see also chapter 9 Warnings and safety precautions)
Warning!
2.2 Connection and switch-on
The basic setting of the current stimulation device has been set for a supply voltage of 230 V. If needed, the
device can be switched over to the following supply voltages:
•
230 V (standard)
•
115 V
Irrespective of the adjusted mains voltage, the device is appropriate for power frequencies of 48 to 62 Hz.
2.2.1 Voltage selection and fuses
1. Unplug the mains plug.
2. The device is protected by 2 fuses on the mains side that are located in a pluggable box at the back of the
device.
3. The adjusted supply can be seen through the small window of the box.
4. With a screw driver the box can be pulled out of the receiver by the small slot.
5. Take the voltage selector out of the receiver and set the correct mains voltage.
Warning!
The scotch tape over the mains socket is easy to remove. Connect the current stimulation device with the
mains cable to a grounded power outlet.