SUSIE
®
S2000 |
User Guide
Scenarios
| 65
“RF module not found”
RF module not connected
Connect the RF module to the USB port
RF module is not identified by
the computer
Close the software and disconnect the RF
module for at least 5 seconds. Plug it back
in and restart the software
CPR is not detected
Compressions are not detected
The current rhythm is healthy
Set rhythm to an unhealthy heart rhythm
Compressions are not
calibrated
Follow the calibration procedure outlined on
page 18
Ventilations are not detected
The respiratory rate is not set
to 0 per min
Set respiration rate to zero
Ventilations are not calibrated
Follow the calibration procedure outlined on
page 18