SUSIE® S2000

Prehospital and Nursing Care

User Guide

User Guide 17.10.1

© Gaumard Scientific Company, 2017

All Rights Reserved

www.Gaumard.com

SUSIE

®

  S2000  is  an  interactive  educational  system 

developed to assist a certified instructor. It is not a substitute 

for  a  comprehensive  understanding  of  the  subject  matter 

and not intended for clinical decision making.

Summary of Contents for SUSIE S2000

Page 1: ...Scientific Company 2017 All Rights Reserved www Gaumard com SUSIE S2000 is an interactive educational system developed to assist a certified instructor It is not a substitute for a comprehensive unde...

Page 2: ......

Page 3: ...ding 2 Procedures 2 Cleaning 3 Stoma and Breast Examination Insert Guidelines Warnings and Maintenance 3 Electrical Therapy Warnings 4 2 1 Features 5 Airway 5 Spontaneous Breathing 5 Cardiac 5 Circula...

Page 4: ...Electrical Therapy 20 Palpable Pulses 25 Programmable Blood Pressure 25 Oxygen Saturation Monitor 27 Intravenous Extremities 29 Finger Stick Bleed 32 4 4 Neurologic 33 Reactive Eyes 33 Seizures 36 4...

Page 5: ...tes 57 5 3 Status Details Controls 58 Changing Vital Signs 59 Applying Changes 60 Creating Palette Items 61 5 4 Streaming Audio 63 6 Appendix 64 6 1 Troubleshooting 64 Communication Power Issues 64 CP...

Page 6: ...ship Electronic transmission is warranted to be free from defects at the moment of transmission Your sole and exclusive remedy and Gaumard s sole liability is to replace the defective media or to repe...

Page 7: ...al Ball point pens ink and markers permanently stain the skin Do not wrap this or any other Gaumard product in newsprint Marks made with ballpoint pens ink or marker cannot be removed Replacement part...

Page 8: ...tems please contact customer service WARNING Vein tubing contains latex which may cause allergic reactions Users allergic or sensitive to latex should avoid contact Discontinue use of this product and...

Page 9: ...clean hands Always palpate the breast using the fatty pads of the middle three fingers Prevent items from resting or pressing against the stomas breasts as indentations will form on the pressure point...

Page 10: ...d arcing Make sure the simulator is not in contact with any electrically conductive surfaces Use the simulator only in a well ventilated area free of all flammable gases NEVER attempt to service or mo...

Page 11: ...ble breathing patterns Accommodates assisted ventilation including bag valve mask and mechanical support with chest rise Cardiac Multiple heart sounds Chest compressions are measured and logged Electr...

Page 12: ...theterization and irrigation Enema exercises BSE Techniques Interchangeable left breasts Right breast contains 8 10 16 and 20 mm lumps Other One year limited warranty 2 2 Terminology Facilitator The p...

Page 13: ...simulator on a patient table capable of supporting the weight of a real adult patient WARNING To avoid damage to the simulator please store and ship it in the clear poly bag provided Leg Assembly Fol...

Page 14: ...nation inserts are shipped in a protective case There are two right breasts labeled R1 R2 and five left breasts L3 L4 L5 L6 L7 WARNING Always store the breast inserts inside the protective case after...

Page 15: ...red under warranty 4 Carefully place the breast inserts beneath the chest skin 5 Secure the chest skin by first attaching the VELCRO located on both shoulders 6 Adjust the skin around the power port l...

Page 16: ...together 3 Unscrew the waist knob and lift the chest skin slightly 4 Tuck the battery wiring underneath the skin 5 Screw the waist knobs back in place Plug the charger into the port and let her charge...

Page 17: ...ed battery It is good practice to re charge the battery at the end of every simulation session In addition re charged the battery at least once every 2 months even if the simulator is not being used p...

Page 18: ...SUSIE S2000 User Guide 12 Equipment Setup 4 Select Susie S2000 and click START Susie S2000 will establish communication within 1 minute...

Page 19: ...bes The airway has anatomical landmarks and can be intubated orally or nasally If the endotracheal tube is inserted too deep the left lung is automatically disabled realistically demonstrating right m...

Page 20: ...ecommended Device Size Intubation Blade size Miller 4 or MAC 3 5 LMA Size 4 Nasal Intubation 8 mm catheter Oral Intubation ETT 7 0 or 7 5 Airway Complication Use the software controls to enable the ai...

Page 21: ...hea to be pierced during a simulation with a real device to observe chest rise Lubricate the tracheostomy tube before inserting To replace the tracheostomy tape that has be punctured follow the steps...

Page 22: ...ttern and inspiration percentage using the software controls The breathing patterns are synchronized with the lung sounds and chest rise Lung Sounds Left and right lung sounds are available normal whe...

Page 23: ...NI software WARNING Do not perform mouth to mouth ventilation Doing so may lead to molding of the airway The airway itself cannot be sanitized or cleaned 4 3 Circulation Central Cyanosis Susie display...

Page 24: ...ng chest compressions Chest compressions are measured and logged by the UNI software Compression Calibration To calibrate the compression feature using UNI tablet 1 Click the gear icon on the upper ri...

Page 25: ...g SUSIE S2000 19 4 Select Normal Calibration and click Next 5 Click Start 6 Perform 1 compression and the oval will turn green 7 Perform a total of 5 compressions then click Finish 8 Finally click Sav...

Page 26: ...d Green Doing so will not create a fire hazard nor is there risk of shock to the provider but internal damage to the simulator will result This type of damage is considered improper use and it is NOT...

Page 27: ...ormal sinus rhythm ECG Snap Option Purchase the ECG Snap chest skin that connects directly to the ECG leads No ECG electrodes are necessary Defibrillation Snaps Option The defibrillation option ordere...

Page 28: ...uide 22 Using SUSIE S2000 2 Remove the waist knobs 3 Gently lift the chest skin 4 Remove the breast and set aside 5 Carefully pull out the Defibrillator Snap harness 6 Press the clip down and disconne...

Page 29: ...0 23 7 Connect the side labeled Defib Snap Vest to the chest skin connector 8 Connect the side labeled Defib Snap to the ECG module 9 Tuck the cables into the cavity 10 Fit the breast back in place 11...

Page 30: ...nap connector covers at the sternum and apex sites 2 Connect the RED adapter to the Apex Snap 3 Connect the BLACK adapter to the Sternum Snap Procedure Maximum Joules Defibrillation 360 Joules Snap Ca...

Page 31: ...hypotension Proper chest compressions result in palpable carotid pulses The simulator s pulses are dependent on blood pressure heart rate and rhythm Programmable Blood Pressure Programmable blood pre...

Page 32: ...it would be placed as instructed above To calibrate the blood pressure feature using UNI tablet 1 Tap the gear icon on the upper right corner of the screen 2 Select Setup then Calibration 3 Click Bloo...

Page 33: ...eft index finger Before using the oxygen saturation feature for the first time calibrate the simulator to work with the oxygen saturation monitor to be used during the exercise Oxygen Saturation Calib...

Page 34: ...USIE S2000 4 Select Setup then Calibration 5 Click Oxygen Saturation and then click Next 6 Click OK in the information window 7 Select Simulator s LEFT index finger then click Next 8 Turn on the oxyge...

Page 35: ...Wait about 20 30 seconds after making an adjustment 11 Click OK when the monitor matches the UNI value 12 Click Next to continue 13 Follow the prompts and then click Finish Intravenous Extremities SUS...

Page 36: ...ard s provided simulated blood Any other simulated blood brand containing sugar or any additive may cause blockage and or interruption of the vasculature system Instructions for Use 1 First locate the...

Page 37: ...ollapse the veins Disconnect the syringe tube from the arm port while maintaining suction The port will seal and the veins will remain collapsed Cleaning the Vasculature Clean and dry the forearm vasc...

Page 38: ...d samples WARNING Use only Gaumard s provided simulated blood to fill the internal reservoirs Any other simulated blood brand containing sugar or any additive may cause blockage and or interruption of...

Page 39: ...nger stick fill syringe to the black port 4 Inject up to 2 cc s of simulated blood or fluid 5 Remove the fill syringe 6 Enable the feature in the UNI software 4 4 Neurologic Reactive Eyes Susie is equ...

Page 40: ...ilation Pupil Calibration The eye reaction is factory calibrated Recalibrate the pupil reaction for the current room lighting only if needed 1 Tap the gear icon on the upper right corner of the screen...

Page 41: ...2000 35 5 Click Ambient Light to calibrate to the current ambient light 6 Cover Susie s eyes and select Low Intensity Light 7 To adjust the sensitivity to light click Increase or Decrease 8 Click Fini...

Page 42: ...ped to assist health professionals in teaching the processes and skills required to perform both breast self examinations and clinical identification of pathologic conditions Breast Pathology R1 Norma...

Page 43: ...ease review the Care and Maintenance guidelines on page 3 Bowel Sounds SUSIE S2000 produces audible bowel sounds Sound types and volumes are controlled from the tablet Auscultate using a standard stet...

Page 44: ...s for the proper cleaning procedures Nasogastric Feeding The fluid capacity is up to 750ml Before introducing fluids through an NG tube please do the following steps 1 Place the simulator in a 30 incl...

Page 45: ...l Reservoirs for the proper cleaning procedures Draining the Gastric Fluids 1 Turn the simulator ON 2 Locate the stomach intestine fluid access port 3 Connect a syringe to the stomach intestine fluid...

Page 46: ...nd removing intestinal fluids introduced by way of nasogastric feeding Cleaning the Internal Reservoirs To properly clean the internal stomas and stomach reservoirs follow the steps below 1 Inject the...

Page 47: ...voir in the abdominal cavity 8 Secure the abdominal cover back in place Gynecology Package SUSIE S2000 includes a gynecologic package that allows the practice the following gynecologic procedures Vagi...

Page 48: ...rted Uterus 1 Retroverted Uterus 1 IUD Uterus Installed 1 6 8 Week Pregnant Uterus 1 10 12 Week Pregnant Uterus 1 20 Week Pregnant Uterus 1 6 8 Week Pregnant Uterus with short ligaments 1 Normal Cervi...

Page 49: ...To interchange the uterus and or the cervix insert 1 Gently lift the abdominal skin cover 2 Unhook the round ligaments from both sides of the uterus 3 Gently twist the uterus counter clockwise to unf...

Page 50: ...4 white connectors 6 Insert the uterus and gently twist clock wise 7 Hook the round ligaments to the torso 8 Velcro the lower torso skin back in place Urinary Catheterization SUSIE features an intern...

Page 51: ...the male genitalia 1 Remove the female catheter adapter Set the female catheter adapter aside and save for female catheterization procedures 2 Insert the male genitalia urethra tube into the catheter...

Page 52: ...ale or male genitalia Install the female genitalia adapter to prevent leaks around the catheter when performing a female catheterization WARNING Remove all fluids after each use Draining the Bladder 1...

Page 53: ...and drain the contents again Decubitus Ulcers SUSIE S2000 has decubitus ulcers Intramuscular Injection Sites IM sites are located on both deltoids and quadriceps for placement and technique exercises...

Page 54: ...elow to remove the healthy foot and replace it with an ulcerated foot 1 Remove the left healthy foot with the hexagonal wrenches provided 2 Pull out the red pulse line from the foot 3 Separate the hea...

Page 55: ...e 9 Insert the ankle bolts and use the hexagonal wrenches to tighten 4 5 Options Virtual Vital Signs Monitor The optional Gaumard Vital Signs Monitor simulates a vital signs monitor attached to the si...

Page 56: ...ug recognition arm This includes calibration purging draining IV infusion and injecting fluids into the veins or the filling ports Introducing fluids into the drug recognition arm while the simulator...

Page 57: ...t the drain tube and the fill syringe Programming the Syringes The tagged syringes supplied with the Drug Recognition arm must be associated with a virtual medication and a concentration before they a...

Page 58: ...the syringe holder on the simulator s right wrist 5 Place the tagged syringe in the holder 6 Keep the syringe perpendicular to the surface of the forearm 7 Click the Medication drop down menu and sel...

Page 59: ...syringe with fluid 3 Inject a vein on the right forearm 4 The UNI software will log the drug administration The tagged syringe must be close to the arm for the drug recognition module to detect the me...

Page 60: ...the status of the radio link between the tablet s USB RF module and the simulator Full bars indicate excellent communication i e normal operation Back Up Battery Indicator The battery status indicator...

Page 61: ...the simulator to stand by mode and then again to resume 5 2 Quick Launch The UNI interface opens up showing the quick launch page for the scenarios This page is used to easily access the preprogrammed...

Page 62: ...es can be selected at the same time and the list of scenarios on the right will display only the scenarios included in the selected categories Selecting the Scenario Click on one of the scenarios list...

Page 63: ...ds the appropriate scenario and starts it automatically Favorites The Favorites feature allows users to reduce the number of scenarios highlighted to those within the categories that will be used most...

Page 64: ...the simulator s vital signs The individual parameter controls displayed on the details tab provide the simplest method for controlling the simulator s vital signs sounds and features The Status Detai...

Page 65: ...ntrol e g heart rate blood pressure respiratory rate etc Alternatively use the keyboard for manual entry and click the green checkmark to confirm the change To change patterns sounds and rhythms click...

Page 66: ...e simulator s condition until the new settings are submitted using the Apply panel After the list of changes is created click NOW to update the vital signs instantly Alternatively click a trending tim...

Page 67: ...ital sign settings into a single loadable object Use a palette item to update a set of vital signs quickly To create a new palette item set the values for the desired vital signs parameters using the...

Page 68: ...SUSIE S2000 User Guide 62 Working with UNI Select the palette item from the Load Palette Item menu and click Load Click the apply option to submit the changes...

Page 69: ...to the participant s reply Connect the headset to the tablet In UNI locate the streaming icons in the lower right of the screen Select the microphone icon to talk as the simulator s voice Select the h...

Page 70: ...fied Battery is not charged properly Make sure that LED indicator on battery charger goes through the sequence described in its label usually red or orange after plugging it and then green when charge...

Page 71: ...lug it back in and restart the software CPR is not detected Compressions are not detected The current rhythm is healthy Set rhythm to an unhealthy heart rhythm Compressions are not calibrated Follow t...

Page 72: ...e crash passenger in a car that was rear ended Even though the air bags did deploy she had slipped in her seat and was caught in the neck and chest by the seat belt She has significant bruising to her...

Page 73: ...er daughter numerous bouts of vomiting and diarrhea She has a history of high blood pressure hyperlipidemia and arthritis in both hands Additionally is experiencing intermittent episodes of confusion...

Page 74: ...a rapid heart rate She has had a six day history of a productive cough and fever and has a history of pneumonia six months ago and vascular dementia for two years She has been on a course of azithromy...

Page 75: ...t Router to USB power supply Computer can be packaged with router connected 3 Or connect the TPLink to the wall outlet 4 Open the Wireless Network Connection on the Monitor Computer and connect to the...

Page 76: ...click Connect manually 2 Start the UNI control software on the control computer 3 On the UNI menu bar click the Gear Monitors Configuration The simulator Virtual Monitor Setup window is displayed 4 S...

Page 77: ...SUSIE S2000 User Guide Wireless Network 71 11 Click Connect to accept the incoming connection...

Page 78: ...quent purchasers or users of the product acquire the product as is and this limited warranty does not apply This limited warranty applies only to the products manufactured and produced by Gaumard This...

Page 79: ...cess to the simulator for possible troubleshooting as needed Technical Support Email support gaumard com USA 800 882 6655 INT 01 305 971 3790 10 2 General Information Sales and Customer Service E mail...