CODE BLUE® MULTIPURPOSE |
User Guide
22 | Working with the Simulator
4.
Perform a correct compression.
5.
Follow the text cue on the screen to perform
the remaining four compressions. When the
calibration is complete, OMNI® will display DONE.
6.
Press SAVE to complete the
calibration procedure.
NOTE: The simulator is precalibrated to the
current CPR guidelines at time of manufacture.
If the CPR guidelines change, calibrate the
sensors inside the simulator using this option.
4.5 CHEST COMPRESSION CALIBRATION WITH OMNI® 2
1.
Make sure your simulator is connected
to OMNI 2 by following the steps in
Section 3.8 “Connecting OMNI 2”.