TECHNICAL MANUAL

9

GATERD.SE

5. Replacement of remote control

Incorrect remote control is easily changed by releasing the contact / remote control on the left pillar - new 
remote control is mounted in the same way.

NOTE: It is important that contact is locked with the clamp lock to ensure IP X5.

6. Calibration of actuators

The calibration of the actuators (height / lowering only) is normally done fully automatically each time the chair 

reaches its end positions (min height, max tilt).

If the chair still does not go in parallel or if the tilt function does not seem to work, an ACTIVE calibration can be 
done. This is done by holding down <button-slow / up> + <button-slow / down> simultaneously.

After about 15 seconds the chair will be moved to its end positions ie. my height and maximum slope 

backwards. Hold down the buttons until the end positions are reached. Release the buttons and wait 5 
seconds. The chair is now calibrated and can be used again.

 

Calibration should NOT be done with user in chair. Note that all 3 actuators move during the calibration  

 

phase. Otherwise, cancel the calibration immediately and contact Customer Service / GATE (or your  

 

local dealer).

7. Reset service indicator

The remote control has a service indicator for periodic maintenance reminder (min once a year). The service 

indicator measures voltage and is activated after 11 months (flashes every 4 seconds). After 12 months the 

indicator will glow steadily.
The indicator is reset by holding down buttons on the remote control (top line, <fast up>, <fast down>) for 30 
seconds.

Summary of Contents for EASY HighBack

Page 1: ...A DIRECT HEALTHCARE GROUP COMPANY G ATERD S E EASY HighBack Teknisk manual ...

Page 2: ...EASY HIGHBACK GATERD SE 2 ...

Page 3: ... in the high back version HB is a development of the conventional concept using the back strap neck support already available This hygiene chair is fitted with a high padded water repellent back complete with neck support available in two versions The high backrest provides soft yet stable support and its lower section is fitted with a pliant lumbar support for additional comfort EASY HB is design...

Page 4: ...t tube left right 11 58 296 1 Footrest 12 58 281 1 Bottom frame 13 57 049 RF Wheel with bolt fully braked 13 57 040 RF Wheel with bolt directional lock and fully braked 14 58 360 Cover 15 57 793 End plug 16 58 253 Battery Pb 5 0 Ah 17 58 291 Cover control box 18 58 293 Bellows silicone 19 58 282 Cap tilt actuator 20 58 252 Actuator TA16 21 58 290 Cover plate left 21 58 289 Cover plate right incl h...

Page 5: ...TECHNICAL MANUAL 5 GATERD SE ...

Page 6: ...213 Splashback 56 214 Armrest lock M8 56 216 Infill 56 217 Waist belt thigh belt 56 218 Slides 56 220 EASY Heel strap kit pair Easy 56 221 HB Width kit 50mm Easy HB 56 227 L Calf rest left 56 227 R Calf rest right 56 229 Chest belt 56 232 Plug soft seat 56 234 Comfort seat back opening green Easy 56 260 L Footrest left 56 260 R Footrest right ...

Page 7: ... with a plastic bag or the like that can withstand the wa ter pressure of the washing machine Place remote control control box or connector female so that running water cannot penetrate Seat can be removed and cleaned separately or washed in machine autoclave max 85 degrees The pillars cleaned with commonly used cleaning disinfectants in health care alcohol based The pillars should be cleaned with...

Page 8: ...y are whole these must be replaced if they are damaged 3 The seat soft seat is complete and without cracks No sharp edges may be present 4 Footrest plates are whole and possible to adjust in the intended way Make sure clips are fully and securely fastened The heel straps should be located on the back of each footrest plate 5 Wheels Do not roll slowly or sit loose Also check that the brake is worki...

Page 9: ...n simultaneously After about 15 seconds the chair will be moved to its end positions ie my height and maximum slope backwards Hold down the buttons until the end positions are reached Release the buttons and wait 5 seconds The chair is now calibrated and can be used again Calibration should NOT be done with user in chair Note that all 3 actuators move during the calibration phase Otherwise cancel ...

Page 10: ...y Off button see below Wait a minute To activate the battery press its On button see below Make sure the battery indicator light is on or flashing 1a Indicator shows green Calibrate actuator see page 7 The indicator light is on flashing Replace the remote control and charge the battery as instructed 1b Indicator shows red Charge the battery The indicator light is not on flashing Change battery 1c ...

Page 11: ...arranty We normally provide a 2 year warranty on manufacturing defects does not apply to wear details or battery The battery is given a 6 month warranty on manufacturing defects For more information or questions please contact Gate Rehab Development AB Industrigatan 2 S 333 75 Reftele tel 0371 318 00 info gaterd se www gaterd se 1 2 3 4 5 6 7 A DIRECT HEALTHCARE GROUP COMPANY ...

Page 12: ... Lift 100kg pelare 150 kg Stroke 400 mm 75 mm Installation dimensions 377 mm 198 mm Protection class N A IP66W Protection class pillars IPX5 Speed without load 32 mm s 11 5 mm s Speed full load 19 mm s 7 mm s Blocking when lifting 100 kg 150 kg Operating temperature 5 45 C 5 45 C Battery TBB6 Voltage 24 V Capacity 5 Ah Type Pb Protection class IP66W Battery charger TP9 Voltage in 100 240V 50 60 Hz...

Page 13: ...by removing the protective cap 4 screws Control box with battery with removed cover Actuators 2x up down and 1x tilt The actuators are connected directly to the control box as below Remote control and battery are connected via spindle cable to control box as below A B 1 1 5 7 6 7 2 3 4 ...

Page 14: ...e Hygiene Chair s electrical components or cabling with original parts from supplier Failure to do so may affect the product s EMC approval Avoid placing using the chair less than 0 35m from other medical equipment and or mobile phones or other transmitting equipment This is because there is then a risk that the main functions of the product will not work as intended or that our product may interf...

Page 15: ... differential mode n a for common mode Mains power quality should be that of a typical commercial or hospital environment Voltage dips short interruptions and voltage variations on power supply input lines IEC 61000 4 11 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for 5 cycles 70 UT 30 dip in UT for 25 cycles 5 UT 95 dip in UT for 5 sec 5 UT 95 dip in UT for 0 5 cycle 40 UT 60 dip in UT for...

Page 16: ...s than the compliance level in each frequency range b Interference may occur in the vicinity of equipment marked with the following symbol NOTE 1 At 80MHz and 800MHz the higher frequency range applies NOTE 2 These guidelines may not apply in all situations Electromagnetic propagation is affected by absorption and reflected from structures objects and people Field strengths from fixed transmitters ...

Page 17: ...TECHNICAL MANUAL 17 GATERD SE 15 Dimensions All dimensions in millimeters and with 125 mm wheels When using Double wheels 100mm width over wheels increases by 30mm ...

Page 18: ...EASY HIGHBACK GATERD SE 18 ...

Page 19: ...TECHNICAL MANUAL 19 GATERD SE ...

Page 20: ...GATER D SE Gate Rehab Development AB Industrigatan 2 Reftele Sverige S 330 21 T 0371 318 00 F 0371 318 01 E info gaterd se LIT 00113 Issue 1 Date January 2020 A DIRECT HEALTHCARE GROUP COMPANY ...

Reviews: