background image

View history data 

Press the MENU, choose “view history data”. Then at the top is the record total number, and the reminder 

“PLEASE SELECT”. Now, if you press MINUS, then the record data will be from the first record to the second in turn. If 

the record number is above 10, and you want to check the specific number, you press the MENU again, then you can 

input the exact number and check that record(press the MENU to move the cursor, and press the MINUS and PLUS to 

change the number), press ON/OFF to confirm and then you will see the detailed record.  

In the detailed record interface, at the top are the record total number and current number. In the middle are the date 

and time, and the concentration is at the bottom. 

Erase history data 

To make sure the detector work perfectly and you can check the history data quickly, we suggest you to erase the 

history data regularly. Press MENU, choose the “Erase ex_flash”, password is needed (the password is 9999), then 

press ON/OFF, it shows that “erasing”, and when finished, it will turn back to the detecting interface.  

Target Point Calibration 

Note: Calibration needs to be done by professionals and with necessary accessories and standard gas.

 

Prepare the standard gas before starting the target point calibration. In detection mode, press ”MENU”, select “calibrate 

span”, press ON/OFF button to enter. Then you will need to enter password. (If you need password, please contact the 

manufacturer to get it) 

After entering the password successfully, press ON/OFF to enter the target point calibration interface, as shown in the 

picture. The interface will display “concentration”, “set value. The concentration is the result got from the calculation of 

several internal calibration parameters. The process of calibration is the process to recalculate the internal parameters 

of the device. The set value is the value need to be entered manually; it should be the concentration of the standard 

gas to be calibrated. Select the parameter under the current interface, modify its value with “PLUS” and “MINUS” 

button.  

The calibration process is as below: 

1.

Connect the detector with standard gas cylinder by hose.

2.

Enter the target point calibration interface, set the “SET value” as the standard

 

gas

concentration.

3.

Release the standard gas to the detector in a flow rate of 500mL/min. The value

 

of

“concentration” will become larger gradually, and become stable after

 

30seconds.

4.

When the value of “concentration” is stable, press the “ON/OFF” button, the

 

value of

“concentration” will be same with or almost same with the “Set value”.

 

The calibration

is finished.

5.

Cut off the standard gas supply. Press “ESC” to go back to detection interface.

www.gasdog.com

[email protected]

+86 773-280-9002

9

Summary of Contents for GD200

Page 1: ...the detector Unique design small size and easy to carry with Special high intensity engineering plastics housing suitable for different kinds of situation and environment We dedicate to provide you th...

Page 2: ...bration functions makes the detection more accurate and reliable One press to restore factory setting free from the bother of mis operation With temperature and pressure compensation this is able to r...

Page 3: ...dapter Instruction Qualification certificate A charging adapter Accessories Configuration List Application Petrochemical Chemcial Industry Municipal Engineering Utilities Agricultral Environmental Pro...

Page 4: ...10ppm 0 2000ppm 0 100 VOL 0 25 VOL 0 2000ppm 0 1000ppm 0 200ppm 0 100ppm 0 100ppm 0 250ppm 0 100ppm 0 50ppm 0 100ppm 0 100 VOL 0 1 LEL 0 1 VOL 0 01 VOL 0 1 VOL 0 1ppm 1ppm 0 01ppm 0 1ppm 0 01ppm 0 1p...

Page 5: ...ng Gardening Interior Decoration and Renovation Dyestuff Papermaking Pharmacy Health Care Foodstuff Antiseptic Disinfection Chemical Fertilizer Resin Adhesive Pesticide Raw Material Sample Technologic...

Page 6: ...or three seconds the detector will be turned on and the pump will start to work The screen displays the following interface successively Brand and Logo main parameter interface initialing countdown Th...

Page 7: ...ding back to zero before powering it off and do the zero point calibration next time before using it If the reading maintained at full scale user should send the device back to manufacturer or agent f...

Page 8: ...and press the PLUS or MINUS to change the value After the setting is done press ON OFF button to save the settings Low Battery Alarm When the battery is low there will be sound light low battery alar...

Page 9: ...on In detection mode press MENU select calibrate span press ON OFF button to enter Then you will need to enter password If you need password please contact the manufacturer to get it After entering th...

Reviews: