READ FIRST BEFORE OPERATION
The
Multi Gas Clip Simple Plus
is a personal safety device designed to detect
the presence of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2)
and combustible gases (LEL). Before operation, ensure that you have been
properly trained on the use of the detector and the appropriate actions to take
in the event of an alarm condition.
Turning On the Detector
Once activated, the detector will run continuously for 36 months. To activate,
press and briefly hold down the power/menu button. Once activated, the
detector will begin to display Gas Readings (5) immediately. Each Gas
Identifier (6) will show a chasing “0” for the Gas Reading (5) while it is
stabilizing and self-testing. Once all Gas Identifiers (6) have completed the
warm up and stabilization sequence, <65 seconds, the detector is ready to
detect all applicable gases. Display will show:
Bump Testing and Calibrating
These operations can be performed manually or automatically using the
MGC Simple Clip Dock
. Refer to the
Multi Gas Clip Simple Plus User’s
Manual
for instructions.
Auto Zeroing
Continuously hold down the power/menu button to auto zero the detector.
Display will show “ZERO” and count down from 3 to 1, followed by “ZERO OK”.
Release button when display shows “ZERO OK”. The display will change to
“AUTO ZERO”. Wait a few seconds, then press the button briefly when display
shows “APPLY GAS”. Detector has been successfully auto zeroed.
When a Gas Reading (5) exceeds its alarm threshold, the detector will go into
visual, vibrating and audible alarm. On the display, the icon will appear
beside the Alarm Condition (1) and the associated Gas Identifier (6) will flash
off and on continuously.
Warning
- Users must familiarize themselves with the icons in both
alarm and non-alarm states.
Warning
- If the display is missing icons, or cannot be clearly read,
discontinue use and contact Gas Clip Technologies.
Warning -
The IR LEL sensor will not detect hydrogen (H2) or acetylene
(C2H2) gases however, due to their cross-sensitive nature, the CO sensor
will go into alarm well below 10% LEL.
Display Components
Factory Default Alarms
Icon
Description
1
Alarm Condition
2
Life Remaining
3
Calibration/Test Mode
4
Battery Indicator
5
Gas Readings
6
Gas Identifiers
Sensor
LOW
HIGH
TWA
STEL
H
2
S
10 ppm
15 ppm
10 ppm
15 ppm
CO
35 ppm
200 ppm
35 ppm
50 ppm
O
2
19.5%
23.5%
LEL
10%
20%
Warning
- Before each use,
check that all sensor and
alarm ports are clear of any
debris or obstructions.
Stabilizing
Fully operational
6
5
1
3
2
4
Warning
- DO NOT auto zero
the detector in a combustible
atmosphere.
Warning
- DO NOT auto zero
in temperatures above 104°F
(40°C) or below 32°F (0°C).