READ FIRST BEFORE OPERATION

The Multi Gas Clip is a personal safety device designed to detect the presence 

of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2) and combustible 

gases (LEL). Before operation, ensure that you have been properly trained on 

the use of the detector and the appropriate actions to take in the event of an 

alarm condition.

Turning On the Detector 

To activate the detector, press and briefly hold down the power/menu button.

Activation tone will sound. Once activated, the detector will begin to display

Gas Readings immediately. Each Gas Identifier will show a chasing “0” for the

Gas Reading while it is stabilizing and self-testing. Once all Gas Identifiers

have completed the warm up and stabilization sequence, < 65 seconds, the

detector is ready to detect all applicable gases. Display will show:

Turning Off the Detector 

Press and hold down the power/menu button until the “OFF” countdown is 

complete. During the countdown, the detector will beep and flash 4 times 

then turn off. Display will show:

Manually Calibrating the Detector

Continuously hold down the power/menu button through the “OFF” 

countdown and the “CAL ” countdown that follows it. During these two 

countdowns, the detector will beep and flash 8 times with a pause between 

countdowns. Release the button when display shows “AUTO ZERO”. Display 

then shows “APPLY GAS”. Snap on the MGC calibration cap connected to a 

hose, regulator and gas cylinder and apply gas. Calibration is complete when 

the next calibration due date is displayed and the detector goes into alarm. 

Turn gas off and remove calibration cap from detector. Display will show:

When a Gas Reading (4) exceeds its alarm threshold, the detector will go into 
visual, vibrating and audible alarm. On the display, the   icon will appear 

beside the Alarm Condition (1) and the associated Gas Identifier (5) will flash 

off and on continuously.

 Warning

 - Users must familiarize themselves with the icons in both alarm 

and non-alarm states.

 Warning

 - If the display is missing icons, or cannot be clearly read, 

discontinue use and contact Gas Clip Technologies.

 Warning

 - The IR LEL sensor will not detect hydrogen (H2) or acetylene 

(C2H2) gases however, due to their cross-sensitive nature, the CO sensor will 

go into alarm well below 10% LEL.

Display Components

Factory Default Alarms 

Icon

Description

1

Alarm Condition

2

Calibration/Test Mode

3

Battery Charge Level

4

Gas Readings

5

Gas Identifiers

Sensor

LOW

HIGH

TWA

STEL

H

2

S

10 ppm

15 ppm

10 ppm

15 ppm

CO

35 ppm

200 ppm

35 ppm

50 ppm

O

2

19.5%

23.5%

LEL

10%

20%

Warning 

- Before each use, 

check that all sensor and alarm 

ports are clear of any debris or 

obstructions

Stabilizing

Fully operational

Note: Calibration can also be done using the Multi Gas Clip Dock.

5

4

1

2

3