READ FIRST BEFORE OPERATION
The Multi Gas Clip is a personal safety device designed to detect the presence
of hydrogen sulfide (H2S), carbon monoxide (CO), oxygen (O2) and combustible
gases (LEL). Before operation, ensure that you have been properly trained on
the use of the detector and the appropriate actions to take in the event of an
alarm condition.
Turning On the Detector
To activate the detector, press and briefly hold down the power/menu button.
Activation tone will sound. Once activated, the detector will begin to display
Gas Readings immediately. Each Gas Identifier will show a chasing “0” for the
Gas Reading while it is stabilizing and self-testing. Once all Gas Identifiers
have completed the warm up and stabilization sequence, < 65 seconds, the
detector is ready to detect all applicable gases. Display will show:
Turning Off the Detector
Press and hold down the power/menu button until the “OFF” countdown is
complete. During the countdown, the detector will beep and flash 4 times
then turn off. Display will show:
Manually Calibrating the Detector
Continuously hold down the power/menu button through the “OFF”
countdown and the “CAL ” countdown that follows it. During these two
countdowns, the detector will beep and flash 8 times with a pause between
countdowns. Release the button when display shows “AUTO ZERO”. Display
then shows “APPLY GAS”. Snap on the MGC calibration cap connected to a
hose, regulator and gas cylinder and apply gas. Calibration is complete when
the next calibration due date is displayed and the detector goes into alarm.
Turn gas off and remove calibration cap from detector. Display will show:
When a Gas Reading (4) exceeds its alarm threshold, the detector will go into
visual, vibrating and audible alarm. On the display, the icon will appear
beside the Alarm Condition (1) and the associated Gas Identifier (5) will flash
off and on continuously.
Warning
- Users must familiarize themselves with the icons in both alarm
and non-alarm states.
Warning
- If the display is missing icons, or cannot be clearly read,
discontinue use and contact Gas Clip Technologies.
Warning
- The IR LEL sensor will not detect hydrogen (H2) or acetylene
(C2H2) gases however, due to their cross-sensitive nature, the CO sensor will
go into alarm well below 10% LEL.
Display Components
Factory Default Alarms
Icon
Description
1
Alarm Condition
2
Calibration/Test Mode
3
Battery Charge Level
4
Gas Readings
5
Gas Identifiers
Sensor
LOW
HIGH
TWA
STEL
H
2
S
10 ppm
15 ppm
10 ppm
15 ppm
CO
35 ppm
200 ppm
35 ppm
50 ppm
O
2
19.5%
23.5%
LEL
10%
20%
Warning
- Before each use,
check that all sensor and alarm
ports are clear of any debris or
obstructions
Stabilizing
Fully operational
Note: Calibration can also be done using the Multi Gas Clip Dock.