Page 20
7094LP Manual
the length and to make sure that the auto calibration is
working properly.
3. To check the diagnostics, press and hold the button for the
tank to be checked, the display will show the level for that
tank.
4. While continuing to hold down the button for the tank, press
the LPG button. When the display shows ”dIA”, release the
buttons, the display will then change to showing the signal
power diagnostic. This is indicated by a “P” showing on the
left digit, for example “P26” indicates a 26% signal power.
5. The signal power will show for 5 seconds. The display will
then change to showing the sender height. This is indicated
by a small “h” showing on the left digit, for example “h 6”
indicates that the sender has 6 receive segments, which is 9
inches high.
6. After 5 seconds of showing the height, the display will shut
off.