Operation
28
operation manual VCpro 600
6.2.2
Operating mode 2 Automatic
In the operating mode
Automatic
a material separation of a mixture up to maximum 4 com-
ponents can be performed.
PARAMETER
VALUE
UNIT
Detection Gradient
0
mbar / min
∆p Distillation Ramp
0
mbar
∆t Distillation Ramp
00:00:00
hh:mm:ss
Number of Cycles
1
#
Parameter
Values from/to
Resolution
Default
Unit
Detection Gradient
-200...200
1
0
mbar/min
∆p Distillation Ramp
-1000...0
1
0
mbar
∆t Distillation Ramp
01:00:00...23:59:59
01:01:01
1:00:00
hh:mm:ss
Number of Cycles
1...4
1
1
Vent at Finish
Yes / No
No
Maximum RPM Speed
10...100
1
50
%
NOTE
The parameter "RPM" can only be set for the VCpro 602 model.
The separation of the material mixture is done by searching for the boiling point of the re-
spective component via the detection gradient. If the pressure gradient changes with a sud-
den increase or drop when a component begins to boil, then this pressure value will be as-
sumed as start setpoint. The distillation of the components is then done by running the distil-
lation ramp. The user must specify the detection gradient for the boiling and the distillation
ramp for the component separation.