e.bloxx Communication
ASCII PROTOCOL
HB_EBLOXX_COM_E_V12.doc
13
Gantner Instruments Test & Measurement GmbH
3.4. Instruction Parameters
device identification (v...v)
length = x char
<vendor name>
ASCII ("Gantner,")
1..32 char
<model name>
ASCII (<type of module>,)
1..32 char
<hw version>
ASCII ("xy.yy,")
1..8 char
<sw version>
ASCII ("xy.yy,")
1..8 char
device information (s...s)
length = 27 char
<location>
ASCII
20 char
<serial number>
ASCII
6 char
<number of variables>
ASCII
1 char
status information (z...z)
length = 6 char
<variable status>
ASCII
4 char
<module status>
ASCII
2 char
<variable st.>
= K16..K13 K12..K9 K8..K5 K4..K1 =
hex
0XYZ
0 X Y Z
ASCII
"0XYZ"
<module st.>
= M8 M7 M6 M5 M4 M3 M2 M1
=
hex
XY
X
Y
ASCII
"XY"
If the bit Kn in the variable status is set it indicates that an error has occurred in variable n. A variable error is given
when the measuring value is outside of the linearization, e.g. in consequence of a sensor break down or of a short
circuit of transmission.
If the bit Mn in the module status is set it indicates that an error has occurred in the sensor module. Valid is:
M1 = 1: EEPROM
- error
M5 = 1: (currently not occupied)
M2 = 1: FLASH - error
M6 = 1: (currently not occupied)
M3 = 1: ADC - error
M7 = 1: (currently not occupied)
M4 = 1: configuration - error
M8 = 1: (currently not occupied)
variable information (b...b)
length = 29 char
<variable type>
ASCII
1 char
<variable name>
ASCII
20 char
<data format>
ASCII
1 char
<field length>
ASCII
1 char
<decimals>
ASCII
1 char
<unit>
ASCII
4 char
<host input>
ASCII
1 char