4. Operational method 

4.1 The newly bought or long-time not used scales should be connected with power supplier 

before using. Pull out the plug in order to charge the battery, it can be used normally while 

recharging and “            ” will flicker on the display, disappear when full. 

4.2 “            ” switch

Press “        ” when the scale is off, the display will show “ON, Voltage, 

Tare range”, then shows “0”, press this key to turn off. 

4.3 “TARE”: with empty scale platform, the displaying deviates from zero, pressing

TARE

key, 

then the displaying will return to zero; If need to tare, first place the container on the platform, 
after stabilizing, press “TARE” key, the scale displays“0”. Then place the article in container, at   

this time, it shows the net weight of the article, remove the article and the container, the scale 
shows a negative value of container weight. Still press “TARE” to return to“0”. 

4.4 “          ” : the units can be changed between “kg” and “lb” 

4.5 COUNT: sampling of article by

COUNT

  key to enter the counting state 

    4.5.1  Choose  the  quantity  of  samples:  To  precisely  count  the  articles,  firstly  the  quantity  of 

samples for counting should be chosen in according with the weights of the articles. The quantity 
of  samples  can  be  chosen  from“1-10-20-50-100”.  For  a  small  number  of  articles,  the  larger 

quantity of samples, the more precise. 

4.5.2

 

Place one or more samples on the empty pan, the balance will show the total weights of 

samples. Then press 

COUNT

, the displaying will show “1”, at the same time, the displaying 

show unit “pcs”, with indicate that scale is now counting. And all the samples are one unit. Then 

press

     

, the displaying will show between“1-10-20-50-100”. The quantity of articles and 

samples must be the same, and place the similar articles. The total amount of objects is showed, 

press the 

COUNT

again back to the normal situation. 

4.6 Calibration Method 

    Hold  on  “TARE”  and  press  “        ”when  the  scale  is  at  zero  point,  the  display  shows 
“CXXXX-”  (XXX  should  be  the  weight  of  standard  weights,  for  example:  when  displaying 
showing “C 100” expressed the need to place a standard weight 100kg) and enter into calibration 

status, place the calibrating weight on the platform

after stability, the scale displays the weight of 

the calibrating weights and the unit “kg” , till now the calibration is completed and the scale can 
be normally used. Under the calibration status, press “TARE”, it will quit. 
 

PS: Setting and meaning   

Press and hold on “        ” to turn on the scale, displays “ON, voltage” and then enter into setting 

mode, pressing “        ” to change the parameters, pressing “TARE” to change the values of 

parameter. 

  1. C1

……

Sensitivity setting 0 1 2 3 4, the larger number is, the less sensible but the better stable. 

Default setting is 2 or 1. 

C2----Filtering  strength  setting  0  1  2  3,  the  larger  number  is,  the  lower  reaction  speed  but  the 

better stable. Default setting is 2 or 1. 

C3---- Battery save mode, “0”: no saving mode,  “1”: turn off the data display when  stabilization 

for 3 minutes, “2”: turn off the data display when stabilization for 3 minutes, auto power-off after 

1 hour. Default setting is 2.   

C4----For this series no use.. 

C5----bright display mode. 0: darker than 1    ;    1: normal ; 2:lighter than 1 ; Default setting is 1. 