background image

displays “0”. Then place the article in container, at this time, it shows the net weight 
of the article, remove the article and the container, the scale shows a negative value of 
container weight. Still press “TARE” to return to “0”. 
 

3.2.3

 

“LIGHT” backlight key

press “LIGHT”

 key to choose the 

backlight mode, 

“B OFF” backlight always off, “B ON” backlight always on

B AUTO

backlight on 

when the data changes, Off when the data reaches stabilization after 3 seconds. 
 

3.2.4 

“COUNT”  count  key:  sampling  of  article  by  “COUNT”  key  to  enter  the 

counting state 
(1) Choose the quantity of samples: To precisely count the articles, firstly the quantity 
of samples for counting should be chosen in according with the weights of the articles. 
The quantity of samples can be chosen from “1-10-20-50-100”. For a small number of 
articles, the larger quantity of samples, the more precise. 
(2)  Place  one  or  more  samples  on  the  empty  pan,  the  balance  will  show  the  total 
weights of samples. Then press “COUNT”, the displaying will show “1”, at the same 
time,  the  right  of  the  displaying  show  unit  “

pcs

”,  with  indicate  that  scale  is  now  in 

counting  state,  and  all  the  samples  are  one  unit.  Then  press  “UNIT”,  the  displaying 
will show between “1

10

20

50

100”. The quantity of articles and samples must 

be  the  same,  and  place  the  similar  articles.  The  total  amount  of  objects  is  showed, 
press the “COUNT” again back to the normal situation.     

 

3.2.5

 

“UNIT” unit conversion key 

Press “UNIT”, the units can be changed between “kg”. “lb”. (1lb=453.59g). 

 

3.2.6

  “CAL”  Calibration  :  Holding  on  “tare”  key,  then  press  “unit”  until  the 

displaying  window  shows  “C  XXX”  and  the  scale  will  automatic  calibrate,(  XXX 

should  be the weight of standard weights, for example: when displaying showing “C 

20”  express  the  need  to  place  a  standard  weight  20kg),  at  the  same  time,  only  place 

calibrating weights on the platform, after stability, the scale displays the weight of the 

calibrating  weights  and  the  unit  “kg”,  till  now  the  calibration  is  completed  and  the 

scale can be normally used. 

 