3.Operation method

 

3.1

 

General information and setup

 

After  transporting  the  scale  from  a  cold  place  to  warm  room  let  it  rest  for  15  minutes 

without power supply to avoid damages at electronic components due to condensation. 

Place the scale on a solid and level ground. Soft ground (carpet) or a table that bends if a 

heavy weight is placed on the board, can cause wrong results. 

The scale needs a power supply of 6V, either by the AC-adapter or by 4 * 1,5V battery. 

If  the  adapter  is  plugged  in,  the  power  supply  to  the  battery  is  disconnected.  They  can 

remain in the scale, but also if used with rechargeable battery, there is no charge function. 

Connect  the  display  unit  to  the  spiral  cable  that  leads  to  the  plateframe  and fix  the  two 

screws at the plug softly by hand. The metal mount can be used to place the display on a 

wall, or to gain better visibility to the field of view of the LCD when the display is placed on 

a  table.  When  transporting  the  scale,  we  suggest  to  turn  the  wall  bracket  forwards  to 

create a protection of the LCD screen.   

3.2

 

Keypad Functions

 

3.2.1

 

“ON/OFF”  key

holding  down 

“ON/OFF”  key  untill  the  display  shows  “ON”, 

then  the  current  internal  voltage,  scale  range,  at  last, 

displays  from  “F----1”  to 

“F-----9” while performing a self-test. After a view seconds “0,0kg” appears. 
The scale now is immediately able to be used for weighings of heavy materials like 

parcels or persons, for this just place a weight on the platform. But please note that 

after  startup  the  scale  needs  a  view  minutes  to  reach  its  optimal  operation 

temperature and to adjust the electrical components properly. During this time, the 

scale is not as accurate as it can be after about 15 minutes, especially for dosages 

of tiny amounts. If the scale is going to be used several times each day, we suggest 

to start it in the morning and keep it running with the display backlight on AUTO to 

save energy (with deactivated LCD-backlight, the scales energy supply goes down 

to about 0,5W and consumes 

about 1 € of energy if one year never turned off). 

To turn off the scale, hold down the 

“ON/OFF” key till the display shows “OFF”, the 

scale will turn off automatically.   

 

3.2.2

   

“TARE” tare key: If need to tare, first place the empty container on the platform, 

after  stabilizing,  press 

“TARE”  key,  the  scale  displays  “0”.  Then  place  the  article  in 

container, at this time, it shows the net weight of the article. After removing the article and 

the  container,  the  scale  shows  a  negative  value  of  containers  weight.  Press 

“TARE”  to 

return to 

“0”. 

 

3.2.3

 

“LIGHT” backlight key: Press “LIGHT”

 

key to choose the

 

backlight mode. 

“OFF” = 

backlight always off

, “ON” = backlight always on. “AUTO” = backlight only flashes on when 

the  weight  on  the  platform  is  being  changed  and  turns  off  after  the  weight  has  been 

stabilized for several seconds.   

 

Summary of Contents for PSB Series

Page 1: ...50kg 50g 0 5 20kg 50g 20 300kg 100g 0 2 75kg 20g 75 150kg 50g 150 300kg 100g Calibration weight 50kg 100kg 200kg 300kg Dimension of platform 400 400 50mm Power Supply AC DC ADAPTOR Input AC100 240V Output DC6V 300mA Or 4 pieces 1 5V AA batteries 2 Shape and installation of scale 1 ON OFF 2 Battery symbol 3 COUNT Function 4 Light 5 TARE 6 Backside Power Socket Battery Case Data line ...

Page 2: ... The scale now is immediately able to be used for weighings of heavy materials like parcels or persons for this just place a weight on the platform But please note that after startup the scale needs a view minutes to reach its optimal operation temperature and to adjust the electrical components properly During this time the scale is not as accurate as it can be after about 15 minutes especially f...

Page 3: ... NOTE The scale has been calibrated before shipment and comes ready to use If the scale once does no longer work correct and shows wrong results you can use a new calibration to adjust the scale again A calibration can also be done to adjust the scale as good as possible to the local environment conditions at the place of use For a calibration a preset adjustment weight is required see table above...

Page 4: ...ry quiet using location Filter 3 4 Robust but slow for bad environment conditions or as personal scale b C2 filtering strength setting values 0 1 2 3 The higher the value the higher the stability Filter 0 Setting for dispensing every weight change will be shown in the display Filter 1 2 For the most use cases Filter 3 Slow but robust best settings when only bigger items parcels are placed on the s...

Reviews: