G&G GmbH | Displayinformation / Error messages
14
11
Displayinformation / Error messages
F----1
Date over display scope
F----2
A/D-switch disorder
F----3
Function key disorder
F----5
Sensor signal disorder / Load Cell defect
F----L
Lower zero position of sensor signal
F----H
Weighing capacity is exceeded
C----F
Calibration error, instability or not at zero point during calibrating
C----H
Calibration error, items on the pan or at higher zero point during calibrating
E-----
Storage calibration data error, re-calibration is necessary
------
Zero point of the scale disorder / mechanical defect of the scale
Should other error messages occur, switch balance off and then on again. If the error
message remains inform manufacturer.
12
Interface RS 232C
An integrated RS-232-interface enables you to automatically send the weighing data to PC. Relevant codes
and control commands are specified as following.
12.1
Technical Data
8-bit ASCII Code
1 start bit, 8 data bits, 1 stop bits, no parity bit
Bit rates supported include: 600, 1200, 2400, 4800 and 9600 bit/s. (default setting is 600)
Miniature plug is necessary (9 pin D-Sub)
For operation with interface faultless operation is only ensured with the correct G&G – interface
cable (max. 2m)
12.2
Description of the data transfer
Symbols
-/+
minus-sign / for positive numbers this is output as blank
Data value of weight / decimal, depending on weighing value
Unit Units / Pcs.
STP
stop character
CR
Carriage Return
Data format: 14-bits Output (ASCII Code):
-/+
Data
Unit
STP
CR
2 bits
7 bits
3 bits
1 bit
1 bit