G&G GmbH | Unpacking, Setup and Commissioning 

10 

 

1.

 

Take the balance and supplied accessories (a feeder, pan elements) out of the box. 

2.

 

Place the balance on a stable ground not affected by mechanical vibrations and airflows. 

3.

 

Level the balance with foot screws so that the air bubble in water-level is in the middle. 

4.

 

Plug the adapter to the power socket at the back of the balance. 

5.

 

Connect the power adapter to an AC220V ± 10%, 50Hz/60Hz outlet. 

 
Scope of delivery: 

 

Balance 

 

PVC Wind protection (for E150Y, E300Y, E600Y, E1200Y, E300YA, E600YA and E1200YA  available) 

 

Glass Wind protection (for E150Y-C, E300Y-C, E600Y-C, E1200Y-C, E100YB-C und E200YB-C available) 

 

Weighing pan 

 

AC adapter 

 

Operating Instructions 

6.3

 

General power supply 

If the balance is used in a laboratory, it can be powered by the external mains adapter (included). In case 
the balance is used as a mobile weighing instrument, it can be powered by AA batteries (not included) or 
internal accumulator (included). The accumulator is charged via the delivered adapter. 
To extant work time of accumulators (battery), it is possible to turn off display’s backlight and the balance 
automatically. The settings of “Power Save” operation see 

chap. 9.3

 an

chap. 11

. 

Please only use the original supplied mains adapter. Otherwise there is a risk that the Balance be 
impaired. 
If the balance is not used for a long time, remove the battery/accumulator and store them separately. 
Leaking battery liquid could damage the balance. 

6.3.1

 

Battery power supply  

The balances can be supply from mains power witch is delivered with the balance. It is also possible to use 
normal AA batteries. 

6.3.2

 

Mains connection during rechargeable battery operation 

When the balance is connected with the mains adapter, the battery power will be disconnected 
automatically. Therefore, there is no battery charging function under the battery power supply. 

6.3.3

 

Battery exchange 

Remove the battery compartment cover and insert six batteries (AA size) into the battery compartment, 
taking extreme care that the polarities plus (+) and minus (-) is observed.  

6.4

 

Initial Commissioning 

In order to obtain exact results with the electronic balances, your balance must have reached the 
operating temperature (warming up time 30 min.). During this warming up time the balance must be 
connected to the power supply (mains, accumulator or battery). 
The accuracy of the balance depends on the local acceleration of gravity. 
Please read the information carefully in 

chap. “Adjustment”

. 

6.5

 

Adjustment 

As the acceleration value due to gravity is not the same at every location on earth, each balance must be 
coordinated - in compliance with the underlying physical weighing principle - to the existing acceleration 

Summary of Contents for EY Series

Page 1: ...6 0 7 Version 2 0 04 2015 EN Precision balances EY E150Y E300Y E600Y E1200Y E3000Y E6000Y E150Y C E300Y C E600Y C E1200Y C E100YB C E200YB C E300YA E600YA E1200YA E3000YA E6000YA E2000YB E3000YB To e...

Page 2: ...acking Installation 9 6 3 General power supply 10 6 3 1 Battery power supply 10 6 3 2 Mains connection during rechargeable battery operation 10 6 3 3 Battery exchange 10 6 4 Initial Commissioning 10 6...

Page 3: ...12 1 Technical Data 17 12 2 RS232 Pinout 9 pin D Sub 18 12 3 Description of the data transfer 18 12 4 Remote control instructions 18 13 Maintenance and Servicing 19 13 1 Cleaning 19 13 2 Maintenance...

Page 4: ...1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight l...

Page 5: ...ming up time 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz Unze GN Gran dwt Pennyweight lb Pfund LCD...

Page 6: ...ght 1 d 1 d 1 d 1 d 1 d 1 d 1 d Warming up time 30 min 30 min 30 min 30 min 30 min 30 min 30 min Reference piece numbers for piece count 1 10 20 50 100 Weighing units g Gramm ct Karat ozt Feinunze oz...

Page 7: ...or battery accumulator 2 1 Proper Use The balance you purchased is intended to determine the weighing value of material to be weighed It is intended to be used as a non automatic balance i e the mater...

Page 8: ...afety Precautions COUTION Carefully read this operation manual before setup and commissioning Environmental conditions only use indoor max Altitude 2000m Operating temperature 20 C 20 C humidity of ai...

Page 9: ...ou select the right location for your balance Therefore observe the following for the installation site Major display deviations incorrect weighing results may be experienced should electro magnetic f...

Page 10: ...dapter Otherwise there is a risk that the Balance be impaired If the balance is not used for a long time remove the battery accumulator and store them separately Leaking battery liquid could damage th...

Page 11: ...technical measuring Info about adjustment weights can be found on the Internet at www gandg de Operations sequence Observe stable environmental conditions A warming up time 30 min is required for stab...

Page 12: ...Note The PVC wind protection is supplied for model E150Y E300Y E600Y E1200Y E300YA E600YA und E1200YA see chap 1 Technical data Note The Glass wind protection is supplied for model E150Y C E300Y C E6...

Page 13: ...G G GmbH Operating elements 13 8 Operating elements Pic 1 E150Y E300Y E600Y E1200Y E300YA E600YA and E1200YA Pic 2 E3000Y E6000Y E3000YA E6000YA E2000YB and E3000YB...

Page 14: ...lance before the first commissioning Level the balance with the adjustable feet horizontal until the air bubble in the level is in the center of the Level Indicator Give the balance after changing the...

Page 15: ...parts it must know the average part weight i e reference Proceed by putting on a certain number of the parts to be counted The balance determines the total weight and divides it by the number of part...

Page 16: ...re changes must be made only at special conditions of use Overview In the off state press the CAL button and keep it pressed press the ON OFF button turn on the balance Now the balance is accessed to...

Page 17: ...enable Default setting is 1 h Save settings After all points have been configured press CAL button returns the balance automatically into weighing mode i Restore the factory settings In the off state...

Page 18: ...if the value is positive a space is transmitted Data 7 bit Value for weight decimal point Unit 3 bit Weight unit CR 1 bit Carriage Return LF 1 bit Line Feed Example The output value 12 345 kg Output...

Page 19: ...ual vacuum cleaner Remove spillages immediately 13 2 Maintenance and repair The appliance may only be opened by trained service technicians who are authorized by G G Before opening disconnect from pow...

Page 20: ...not at zero Adjustment is no longer correct Great temperature fluctuations Warm up time was ignored Electromagnetic fields static charging choose different location switch off interfering device if p...

Reviews: