Calibration
30
Potentiostat calibration is only required infrequently. Recalibrate your EIS Box under the following
circumstances:
•
It is at least one year since your last calibration.
•
Your instrument has been serviced.
•
You notice breaks or discontinuities in the data curves recorded with your system.
•
The system is being run in an environment that is very different from the previous operating environment.
For example, if the EIS Box was calibrated at 15°C and you are now operating it at 30°C, you should
recalibrate.
Connect the Calibration Cell to the appropriately colored banana plugs on the Cell Cable.