13
Adjusting Pressure
Press the pressure button on the control panel to adjust the level of compression.
the system has four pressure options:
1. No pressure: circulates the ice water only
2. low pressure: inflates and deflates the Wrap from 5-15 mm Hg over approximately 5 minute cycle.
3. Medium pressure: inflates and deflates the Wrap from 5-50 mm Hg over approximately 3 minute cycle.
4. High pressure: inflates and deflates the Wrap from 5-75 mm Hg over approximately 3 minute cycle.
the appropriate pressure icon will flash as the unit cycles pressure. the left-hand side of the icon will
appear and flash as the pressure cycles up, and then the right-hand side will appear and flash as the
pressure cycles down. large Wraps, such as the Full leg Wrap, will have longer pressure cycles than
smaller Wraps, such as the Utility Wrap.
Adjusting Pressure – Wraps
the system will default to the low pressure mode for all Wraps. to adjust the pressure, press
once for medium pressure, twice for high pressure and three times for no pressure.
Changing Pressure during treatment
the pressure cannot be adjusted while the system is in use. to adjust the pressure, press the pause
button, change the pressure to the desired mode, and then press the start button to resume therapy.
temperature
Press the °C/°F button to choose your desired mode. With a full ice box, typically the Control Unit
will provide cold therapy between 35 °F and 50 °F.
Setting treatment time
Each time the power button is used to turn the system on, the hourglass icon will appear and the timer
will automatically set for 15 minutes. If you press start, the system will run for 15 minutes, then stop.
Adjust the length of a timed session by pressing the plus or minus keys to your desired time, from 0-95
minutes, in 5 minute increments. once you‘ve set the desired treatment time, press the start button.
An hourglass icon will appear on the control panel to indicate the timer is running. When the time is
up, the alarm will beep and the system will stop.
Changing timer during treatment
to add or subtract minutes while the timer is running, you must first press pause, then press the plus or
minus buttons. (If you press the timer setting button or the plus or minus buttons while the system is in
use, the system will not respond.) Press the Start / Pause button to resume treatment.
running the System Continuously
If you’d rather run the system continuously, just press the timer button once, the display will change to zero,
and the hourglass icon will disappear. Press start and the system will run and begin counting elapsed time.
the system will run for a maximum of 190 minutes, by counting up to 95 minutes and then back down to zero.
After 190 minutes, the system will stop and remain in pause mode.