GENERAL
IMPORTANT
READ THIS COMPLETE USE GUIDE AND GAME READY GRPRO 2.1 SYSTEM USER
MANUAL INCLUDING INDICATIONS, CONTRAINDICATIONS, CAUTIONS AND
WARNINGS BEFORE USING THIS PRODUCT!
WARNINGS
• THE CRYO CAP IS NOT TO BE USED CONTINUOUSLY FOR GREATER THAN
15 MINUTES PER TREATMENT SESSION
• THIS WRAP IS INTENDED FOR USE WITH LOW-PRESSURE SETTING.
PRESSURE CAN BE TURNED OFF AS DESIRED.
(NOTE: This Wrap is not
designed to deliver compression therapy; low compression is only used to provide
good surface contact.)
• Follow the recommendations of your health care practitioner regarding the
frequency and duration of use.
• Improper placement or prolonged use of the Game Ready System could result
in tissue damage. During the course of therapy, patients should monitor the skin
surrounding the treated region or the digits of the extremities of the treated
limb for any burning, itching, increased swelling, or pain. If any of these signs
present, or any changes in skin appearance occur (such as blisters, increased
redness, discoloration, or other noticeable skin changes), patients are advised to
discontinue use and consult a physician.
• Game Ready Wraps are not sterile; do not place directly against open wounds,
sores, rashes, infections, or stitches. The Wrap may be applied over clothing
or dressing. A layer of clothing between Wrap and skin is recommended for all
patients.
• Game Ready Wraps are available in multiple configurations but are not intended for
all possible physiologic uses. For example, the Ankle Wrap is not designed for use
on the toes and the Back Wrap is not designed for use in the abdominal region.
• To avoid potential damage to the Control Unit, do not use other manufacturers’
Wraps with the Control Unit.
• Use extra caution during the immediate postoperative period, especially when
sedated or on any medication that could alter normal pain sensation. Check the
skin of the treated region frequently and use mid-to-higher (warmer) temperature
range settings or leave more time between treatments, if necessary.
NOTES
Apply Wrap with a uniformly close fit, ensuring there are no kinks which may impede
water flow. Ensure connector hose is placed to prevent the Wrap from folding or
kinking at the hose inlet location of the Wrap.
CARE AND CLEANING
For daily care and to minimize formation of mildew, remove Heat Exchanger from
Sleeve and wipe with dry towel to remove any condensation that may form. Turn
Sleeve inside out and hang both the Sleeve and Heat Exchanger to release excess
moisture. For multi-patient use, if needed, use Sterifab® according to manufacturer’s
instructions to minimize microbe transfer.
For extended care, carefully remove the Heat Exchanger from the Sleeve and turn the
Sleeve inside out. Hand or machine wash the Sleeve in cold water and mild detergent,
or antibacterial soap. Hang to dry. Hand wash the Heat Exchanger with warm water
and mild detergent, do not machine wash or place in a dryer. Hang to dry.
AVERAGE LIFE EXPECTANCY
The life expectancy of Sleeves and Heat Exchangers will vary widely depending on
frequency of use. Please reference chart below to determine when to replace product.
Sleeve
Light Use (Personal) ............................................................ 12 months
Medium Use ....................................................................... 6 months
Heavy Use (Clinic or training facility) .................................. 3 months
Heat Exchanger
Light Use (Personal) ............................................................ 24 months
Medium Use ....................................................................... 18 months
Heavy Use (Clinic or training facility) .................................. 12 months
WARRANTY INFORMATION
Sleeve:
In case of manufacturer defect, Sleeve may be returned within 7 days of
purchase.
Heat Exchanger:
1 year from date of purchase. See warranty card included with
Heat Exchanger.
INSULATION
This Wrap comes with insulation layers for the ears and
forehead. To use the Wrap’s insulation, tuck the foam layers
under the Heat Exchanger in the desired region(s). To use
the Wrap without insulation, and provide more cooling for
the ears and/or forehead, lay the foam on top of the Heat
Exchanger in the desired region(s).
SUPERSEDED
Summary of Contents for 590920-03
Page 7: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...
Page 11: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...
Page 39: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...
Page 47: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...
Page 51: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...
Page 63: ...GAME READY MED4 ELITE Med4 Elite Med4 Elite Med4 Elite 18 Med4 Elite Med4 Elite 18 SUPERSEDED...