WRO 300 H
44
Technical data
HCEN12104 Rev 09.2010
Connection of
external equip-
ment
External connector 8 pins REDEL, yellow.
Remote start
(Not available in the USA)
Max voltage
5 V DC
Max current
100 mA DC
External alarm
Max voltage
24 V AC or DC
Max current
100 mA AC or
DC
External connector 8 pins REDEL, black. USB
Logging interface
Max input volt-
age
±15 V DC
High level min
output voltage
5.0 V DC
Low level max
output voltage
5.0 V DC
Max Current
±5 mA DC
External connector 8 pins REDEL, green.
External equipment
Accessory equipment connected
to the WRO 300 H should comply
with relevant standards (e.g. IEC
60950 for data processing equip-
ment). Furthermore, medical sys-
tems should comply with IEC
60601-1-1. One important issue to
consider, when configuring a sys-
tem, is that the resulting leakage
current for the system does not
exceed acceptable limits.
Sound
Sound power level
Less than 65 dB(A) during normal
operation
Values