Gambro Prismaflo IIS Quick User Manual Download Page 2

306100449_1 © 2013.06. Gambr

o Lundia AB  

“On” or “Heating” mode 

control panel

“On” or “Heating” mode: Increasing/Decreasing the Set 

Temperature

The set temperature can be checked at any time by pressing the 

“Set” Button.

Actions: 

Press the “Set” Button while the control unit is turned on 

(“On” mode) or started (“Heating” mode).

While the display is flashing, you can select any set temperature 
between 33.0°C and 43.0°C in 1°C increments by repeatedly 
pressing the Button “Set’.

When the set temperature is 43.0°C, the selection starts again at 
33.0°C if you press the “Set” Button one more time.

Characteristics:

• 

The set temperature is displayed and flashes for approximately 3 

seconds (e.g. 41.0°C).

• 

After the selection has been made the display will continue to 
flash the set temperature for approximately 3 seconds then revert 

to current temperature.

• 

After changing the set temperature, the audio alarm will beep 
once to confirm the new set temperature.

• 

When the control unit is in Heating Mode, temperature regulation 
is activated with the new set temperature.

• 

The set temperature is retained in memory even after power is 

removed.

Operating the Prismaflo IIS 
blood warmer

1. Plug the main plug 

of the Prismaflo IIS blood warmer into a socket 

(LED “Standby” turns on to show that the control unit is in “Standby” 
mode).

2. Connect the heating sleeve

 to the Prismaflo IIS blood warmer 

according figures below.

3. Press the Button “Standby”

 to switch the Prismaflo IIS blood 

warmer to “On” mode.

4. Check the audible

 and visual indicators and the display: All 

segments of the display and all LEDs are flashing with the audio alarm 
providing a single beep to confirm that the control unit is working 
properly. The “Start” LED flashes and the temperature display shows 
the actual set temperature flashing for about 3 seconds, then the 
current temperature of the heating sleeve is indicated.

5. Press the “Set” Button 

to select another set temperature if needed. 

6. Press the “Start” Button 

twice to switch the Prismaflo IIS blood 

warmer to Heating Mode (“Start” LED is on permanently).

7. Follow the online instructions

 of the Gambro control unit. With 

Prismaflex system, move to next step when reaching “Prepare and 
Connect Solutions” screen.

8. Align the groove 

of the Prismaflo IIS heating sleeve in an upward 

position, making sure the sleeve is not twisted.

9. Place the end

 of the Prismaflex Set blood return line (blue strip) to 

the end of the heating sleeve approximately 4 to 6 cm from the blue 
Luer Lock connector.

 

10. With the thumb,

 insert the Prismaflex Set blood return line into the 

sleeve up to the Gambro control unit cabinet.

11. Put the heating sleeve 

of the Prismaflo IIS blood warmer into the 

support hooks for the heating sleeve with a kink-free inlet of the blood 
return line.

12. Continue with online

 instructions of the Gambro control unit until 

the patient is connected.

Installation

Patient at the right side of Gambro control unit.

If the patient is positioned at the 
right side of the Gambro control 
unit then guide the heating 
profile sleeve directly from the 
right support of the heating 
profile sleeve back to the patient.

Installation

Patient at the left side of Gambro control unit.

If the patient is positioned at the left side of the Gambro control unit, 
the heating sleeve is guided in front of the machine from the right 
support of the heating sleeve to the left support and then back to the 
patient (A, B).

For Customer Support call

 

800-525-2623

14143 Denver West Parkway, Suite 400
Lakewood, CO 80401

Visit us at www.gambro.com 

 

Gambro, Prismaflex and Prismaflo II are registered trademarks of Gambro Group. 

Manufactured by

Stihler Electronic GmbH

Julius-Hoelder-Strasse 36

70597 Stuttgart

Germany

A

B

Reviews: