background image

Prismaflo

®

 IIS: Quick User Guide

Description

The Prismaflo IIS blood warmer is an accessory 

device for use with the Gambro control units 
using Prismaflex system sets.

The Prismaflo IIS blood warmer is used to warm the return 

blood flow.

1. Attachment Device: 

Attaches 

the Prismaflo IIS Blood Warmer to 
the holder which is mounted on the 
backside of the Prismaflex control 
unit

2. Control Panel: 

Control buttons 

and indicators 

3. Flexible Heating Sleeve 

Transfers heat from the internal 

heating wire to the medium to be 
warmed via the inserted blood 
return line

4. Power Supply Cord:

 Supplies 

electricity from the socket to the control unit

5. Connection Cable Heating Sleeve: 

Connection between 

control unit and changeable heating sleeve

6. Adaptor of Heating Sleeve:

 Connection between heating 

sleeve and cable

Control panel

1. “Standby” LED: 

Turns on automatically when the device is in 

“Standby” mode.

2. “Standby” Button: 

Press to switch from “Standby” mode to 

“On” mode.

3. “Start” LED: 

Flashes when the device is in “On” mode 

(heating not yet started). Permanently illuminates when heating 
(“Start” Button is pressed)

4. “Start” Button:

 Press 

to start heating, when the 
device is in “On” mode or 

“Alarm” mode

5. “Alarm” LED: 

Illuminates during an 
alarm (audio alarm 
beeps automatically if an 
alarm condition exists).

6. “Set” Button: 

Press to view actual set temperature.  

Press repeatedly to select temperature (while in “On” & 

“Heating” mode).

7. Temperature display:

 Indicates the current temperature 

of the heating sleeve

8. “Test” Button

 Press to switch to “Test” mode

“On” mode control panel

“On” mode: Warmer is turned on

Actions: 

Press the “Standby” Button to switch the device from 

“Standby” mode to “On” mode.

Characteristics:

•  The LED “Standby” turns off.

•  The device conducts a self-test. All segments of the display 

and all LEDs are flashing. The acoustic alarm signal beeps 
once to confirm that the control unit works properly.

•  The “Start” LED flashes and the display flashes the set 

temperature for approximately 3 seconds.

•  After that the current temperature of the heating sleeve is 

displayed.

•  If current temperature is below 18°C (64.4 °F), the display 

indicates L.

•  If current temperature is above 48°C (118.4 °F), the display 

indicates H.

“Heating” mode control panel

“Heating” mode: Warmer is started

The device can be started 3 seconds after being plugged in.

Actions:

 Press the “Start” Button to switch the device from “On” 

to “Heating” mode to begin heating of the heating sleeve

Characteristics:

• “Start” LED turns on.

•  A self-test is performed. During this test the excessive 

temperature and cable break alarms are activated once to 
check the function of the cut offs.

•  Temperature regulation is activated.

•  The display shows the current temperature of the heating 

sleeve.

•  If current temperature is below 18°C (64.4 °F), the display 

indicates L.

•  If current temperature is above 48°C (118.4 °F), the display 

indicates H.

7

8

6

4

2

5

3

1

Reviews: