Preface
COBE
®
Spectra™ Apheresis System •
Essentials Guide
1-15
Caution:
Ensure lines are attached to correct fluids:
•
AC line (orange spike) to the ACD-A container.
•
The spike on the access saline line to the 0.9% NaCl
container.
Caution:
Ensure the spike on the access saline line and the
needle on the return saline line are attached to the 0.9% NaCl
container.
Caution:
Ensure fluid is flowing into the drip chambers.
Caution:
Completing spillover recovery reduces platelet yield.
Failure to complete spillover recovery may result in WBC
contamination greater than 1 x 10
6
.
Caution:
Failure to complete PIR before Rinseback during LRS or
LRST procedures results in platelet yields lower than those
predicted on the run results screen.
Caution:
Enter complete donor or patient data before entering
Manual mode. Failure to do so may result in erroneous pump flow
rates once you enter Manual mode.
Caution:
The Spectra system does not automatically control
pump flow rates when operating in Manual mode.
In Manual mode, you can enter pump flow rates outside of the
performance specifications described in Chapter 13. Gambro BCT
makes no accuracy claim for values outside the specified
performance range for each pump. The Spectra system DOES
NOT perform run time valve checks in Manual mode.
Entering Manual mode during an LRS or LRST procedure may
result in WBC levels greater than 1 x 10
6
. At the end of the
procedure, a screen appears indicating that the WBC and platelet
content of the product needs to be measured.
Caution:
You must use the Return Flow Controller to perform a
Single-Needle procedure. DO NOT attempt to perform a Single-
Needle procedure without a Return Flow Controller and
appropriate disposable tubing set.
Caution:
Mount the Return Flow Controller high enough to clear
the Spectra system control panel. This prevents damage to either
assembly when swiveling the control panel.
Caution:
To avoid damage to the Return Flow Controller when it
is mounted on a vertical segment of the Spectra system IV pole,
rotate the Return Flow Controller inward and check door
clearances when moving the system. Rotate the unit outward
again for use.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...