RETURN FLOW TOO SLOW!
Alarm LED is red. Pumps stop.
This alarm occurs only during a Single-Needle Run mode.
RETURN FLOW TOO SLOW! Check return.
Increase return flow scale.
CONTINUE
Note:
If you get repeated Single-Needle alarms, see “Repeated
Single-Needle Alarms” on page 12-2.
Probable Cause
Action
Return pressure from the Return Flow Controller too low
Ensure the Single-Needle bag is empty. Set the return flow scale to a value higher than the original. Press
CONTINUE
.
Blood pressure cuff too tight
Ensure the blood pressure cuff is set at 10 to 20 mm Hg.
Needle infiltration
-or-
Occluded access/return vein
Remove the needle and perform a new venipuncture with a new needle. If using the Single-Needle ELP, LRS,
or LRST disposable tubing set, the product will be a 24-hour platelet product unless you attach the new needle
using an approved sterile connection device.
Reposition or replace needle if necessary. If using the Single-Needle ELP, LRS, or LRST disposable tubing set,
the product will be a 24-hour platelet product unless the you attach the new needle using an approved sterile
connection device. Ensure the Single-Needle bag is empty. Press
CONTINUE
.
Blockage in the return line or needle
Check for restrictions such as kinks, clamp, or clotting in the return line or access/return needle. Ensure the
Single-Needle bag is empty. Press
CONTINUE
.
Return Flow Controller not properly set
Ensure the Single-Needle bag has no kinked or twisted lines. Reset the return flow scale to the number the
Spectra system indicated at the beginning of Run mode.
When the Single-Needle bag is empty
, ensure the
Return Flow Controller returns to the proper bag empty position.
Return pressure sensor malfunction
Note the alarm message. Discontinue the procedure. Notify a qualified service technician of the alarm
message.
Summary of Contents for Cobe Spectra
Page 1: ......
Page 7: ...COBE Spectra Apheresis System Essentials Guide 1 1 1 Preface ...
Page 27: ...COBE Spectra Apheresis System Essentials Guide 2 1 2 Introduction ...
Page 75: ...COBE Spectra Apheresis System Essentials Guide 5 1 5 Software Options and Configuration ...
Page 92: ...Software Options and Configuration 5 18 COBE Spectra Apheresis System Essentials Guide ...
Page 93: ...COBE Spectra Apheresis System Essentials Guide 6 1 6 Automatic and Manual Modes ...
Page 98: ...Automatic and Manual Modes 6 6 COBE Spectra Apheresis System Essentials Guide ...
Page 99: ...COBE Spectra Apheresis System Essentials Guide 7 1 7 Alarm Tests ...
Page 103: ...COBE Spectra Apheresis System Essentials Guide 8 1 8 Anticoagulation ...
Page 109: ...COBE Spectra Apheresis System Essentials Guide 9 1 9 Fluid Volumes ...
Page 115: ...COBE Spectra Apheresis System Essentials Guide 10 1 10 General Maintenance ...
Page 120: ...General Maintenance 10 6 COBE Spectra Apheresis System Essentials Guide ...
Page 121: ...COBE Spectra Apheresis System Essentials Guide 11 1 11 Helpful Hints and Recovery Procedures ...
Page 133: ...COBE Spectra Apheresis System Essentials Guide 12 1 12 General Alarms and Troubleshooting ...
Page 197: ...COBE Spectra Apheresis System Essentials Guide 13 1 13 Specifications ...
Page 208: ...Specifications 13 12 COBE Spectra Apheresis System Essentials Guide ...
Page 209: ...COBE Spectra Apheresis System Essentials Guide 14 1 14 Data Input Limits ...
Page 212: ...Data Input Limits 14 4 COBE Spectra Apheresis System Essentials Guide ...
Page 213: ...COBE Spectra Apheresis System Essentials Guide 15 1 15 The Seal Safe System ...
Page 224: ...The Seal Safe System 15 12 COBE Spectra Apheresis System Essentials Guide ...
Page 225: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 226: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 227: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 228: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 229: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 230: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 231: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 232: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 233: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...
Page 234: ...COBE Spectra Apheresis System Essentials Guide This page intentionally left blank ...