HCEN12745 Revision 10.2016
Program version 2.xx
AK 98 Dialysis Machine - Haemodialysis - Double needle
treatment
A:93
12)
Only for single needle treatment:
The
expansion chamber is filled to a suitable
level during priming. Close the clamp
above the expansion chamber. See
Section 5
“Haemodialysis - Single needle
information.
4.2.7.3
Assisted priming
Procedure
1)
Press the
Priming
button.
2)
Press
Priming
.
3)
Select
Assisted priming
.
4)
Press
OK
.
5)
Follow the instructions on the screen.
6)
When the assisted priming procedure is
complete it is possible to perform extra
priming or recirculation. See Section 4.2.8
on page A:94.
CAUTION!
Check carefully that the
venous blood line is correctly
placed in the priming
detector to make sure that
the supervision of alarms is
activated.
7)
When the level in the venous drip chamber
is OK, activate the air detector.
8)
When the assisted priming procedure is
completed, press
Confirm
.
9)
Press
Connect patient
.
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...