AK 98 Dialysis Machine - Attentions
HCEN12745 Revision 10.2016
Program version 2.xx
513 Buzzer failure
513 Buzzer failure
No audible alarm will be issued.
Appears:
When the loudspeaker fails to produce sounds in a alarm or
attention situation.
Machine actions:
None.
Operator action:
Continue treatment and pay attention to visual alarm
indications.
Mute time:
-