AK 98 Dialysis Machine - Technical data and specifications
HCEN12745 Revision 10.2016
Program version 2.xx
Valid under conditions that preventive maintenance and repair have been performed at
least every other year by an authorized technician.
14.6.5
Energy and water consumption
Typical energy consumption and energy delivery to environment and drain at inlet
water temperature of approximately 23 °C given the conditions stated in Table 14-27
on page A:192.
Energy consumption
2.2 kWh
Energy delivery to the environment
1.3 kWh
Energy delivery to the drain
0.9 kWh
Table 14-25. Energy consumption
Water
140 L
A concentrate
2.9 L
B concentrate, liquid
5.0 L
B concentrate, dry powder
0.4 kg
Table 14-26. Consumption of water and dialysis fluid concentrates
Data in Table 14-25
on page A:192 and Table 14-26
“Consumption of water and dialysis fluid concentrates”
on page A:192 have been
obtained given the conditions stated in this table.
Dialysing time
4h plus preparation time and post treatment
operation
Dialysis fluid flow
500 mL/min
Blood flow
300 mL/min
Ultrafiltration flow
0,5 L/h
Dialysis fluid temperature
37 °C
Type of disinfection
Heat disinfection with CLEANCART C
cartridge
Table 14-27. Consumption conditions
14.7
Standards
The machine complies with the following standards:
IEC 60601-1:2005/AMD1:2012/COR1:2014 Medical electrical equipment - Part 1:
General requirements for basic safety and essential performance.
IEC 60601-1-2:2014 Medical electrical equipment - Part 1-2: General requirements for
basic safety and essential performance - Collateral standard: Electromagnetic
compatibility - Requirements and tests.
IEC 60601-1-6:2010/AMD1:2013 Medical electrical equipment - Part 1-6: General
requirements for basic safety and essential performance - Collateral standard:
Usability (+ Amendment 1).
IEC 60601-1-8:2006 + AM1:2012 Medical electrical equipment -- Part 1-8: General
requirements for basic safety and essential performance - Collateral Standard:
General requirements, tests and guidance for alarm systems in medical electrical
equipment and medical electrical systems.
IEC 60601-1-11:2015 Medical electrical equipment - Part 1-11: General requirements
for basic safety and essential performance - Collateral Standard: Requirements for
Summary of Contents for AK 98
Page 1: ...AK 98 Dialysis Machine Operator s Manual Program version 2 xx Order number MHCEN12745 10 16 ...
Page 45: ...This page is intentionally left blank This page is intentionally left blank ...
Page 106: ...This page is intentionally left blank This page is intentionally left blank ...
Page 118: ...This page is intentionally left blank This page is intentionally left blank ...
Page 128: ...This page is intentionally left blank This page is intentionally left blank ...
Page 144: ...This page is intentionally left blank This page is intentionally left blank ...
Page 166: ...This page is intentionally left blank This page is intentionally left blank ...
Page 196: ...This page is intentionally left blank This page is intentionally left blank ...
Page 198: ...This page is intentionally left blank This page is intentionally left blank ...
Page 208: ...This page is intentionally left blank This page is intentionally left blank ...
Page 268: ...This page is intentionally left blank This page is intentionally left blank ...