How to operate.

Installation & maintenance

PR

ODUC

T M

ANU

AL

RAIR-01 Version 6. Approved 23/02/2022

from the makers of

Summary of Contents for Clinell Rediair RAIR-AUS

Page 1: ...How to operate Installation maintenance P R O D U C T M A N U A L RAIR 01 Version 6 Approved 23 02 2022 from the makers of...

Page 2: ...PRODUCT CONTENTS Rediair decontamination device Pre filter Rediair decontamination device pre filter main combination filter HEPA 14 instruction manual Main combination filter HEPA 14...

Page 3: ...y precautions 5 How to use Rediair Setup 6 LCD touch control panel 8 Disposal cleaning and maintenance External surface cleaning 11 Filter changing and cleaning 11 Cleaning a pre filter 12 Cleaning an...

Page 4: ...ir quality indicator strip monitors a room s air quality and changes depending on the current air quality Rediair will change its internal fan speeds to account for the CADR required to decontaminate...

Page 5: ...the filter is saturated the filter must be replaced GAMA Healthcare recommends replacement every six months In case of strong odour pollution a more frequent change is advisable 1 3 Highly efficient...

Page 6: ...ator Consult instructions for use AU Regulatory Compliance Mark CE marked product European Conformity mark UK Conformity Assessed Do not dispose of in general waste Waste Electrical and Electronic Equ...

Page 7: ...4 Maximum altitude Double insulated device Handle with care Keep dry Temperature limitation Atmospheric pressure limitation Humidity limitation...

Page 8: ...mes wet discontinue use until dry and then test the device for proper operation before reuse Please speak to your local Rediair customer service representative for more information about the warnings...

Page 9: ...the two clips at the top of the filter and remove both sides Remove the main filters by pulling on the tabs both sides Remove the blue tape on the external covers of the device Before getting started...

Page 10: ...re facing outwards Remove the plastic wrap from the filters Connect the power supply Re install the pre filters both sides Re install the external covers both sides Note Ensure external covers are fit...

Page 11: ...o provide maximum decontamination turbo mode allows the Rediair fan to operate at its highest speeds 3 3 3 Night mode Rediair will operate at its quietest level All lights and displays are turned off...

Page 12: ...ce the main filters If the filter indicator flashes red intermittently it is time to clean the pre filters 3 6 1 Working time display Press the filter indicator button to view working time of the curr...

Page 13: ...10 LED AIR QUALITY INDICATOR Very good Not good Good Bad 3 7...

Page 14: ...by the filter indicator button located on the digital display Pre filters should be cleaned regularly at least once a month When the filter indicator changes to red after 3000 hours approximately twi...

Page 15: ...a HEPA filter if available The pre filter can also be rinsed with clean water but must be completely dry before being used again as any remaining moisture could potentially cause damage to the HEPA 1...

Page 16: ...ters both sides Note Ensure external covers are fitted correctly or the Rediair will not power on g m3 PM2 5 1h 2h 4h 8h 12h AUTO FILTER TIMER LOCK POWER MODE SPEED To clean Rediair s air outlet lift...

Page 17: ...ies Also if the device has been used in polluted air situations the filter must be packed into an airtight bag and disposed of per local waste guidance policies In the case of contamination with virus...

Page 18: ...dispose of them both sides g m3 PM2 5 1h 2h 4h 8h 12h AUTO POWER MODE SPEED Switch off device and disconnect the power supply g m3 PM2 5 1h 2h 4h 8h 12h AUTO FILTER TIMER LOCK POWER MODE SPEED FILTER...

Page 19: ...h control panel Install the new pre filters both sides Re install the external covers both sides Note Ensure external covers are fitted correctly or the Rediair will not power on Connect the power sup...

Page 20: ...bon HEPA filter FILTER CLASS HEPA ELECTRICAL CONNECTION V HZ 230 50 POWER W 85 NOISE LEVEL MIN DB A 3 M 22 NOISE LEVEL MAX DB A 3 M 66 CABLE LENGTH M 2 IP CLASS 20 LENGTH WIDTH HEIGHT MM 400 400 669 W...

Page 21: ...ng agents may be used on external surfaces only CHEMICAL FLUID CONCENTRATION CLINELL UNIVERSAL WIPES N A CHLORINE 1000ppm HYDROGEN PEROXIDE WIPES 7 5 Use only the cleaning agents recommended Damage oc...

Page 22: ...EC 60335 2 65 EN 62233 IEC 62321 NAME OF COMPANY ADDRESS DISTRIBUTOR GAMA Healthcare Ltd The Maylands Building Maylands Avenue Hemel Hempstead Industrial Estate Hemel Hempstead HP2 7TG United Kingdom...

Page 23: ...an 60 countries Our Clinell Universal Wipes are the NHS number 1 most used disinfectant wipe trusted by 9 out of 10 NHS hospitals to help safeguard patients and healthcare workers from the risk of inf...

Page 24: ...23 02 2022 JBN220094 GAMA Healthcare Ltd The Maylands Building Maylands Avenue Hemel Hempstead Hertfordshire HP2 7TG UK Tel 44 0 20 7993 0030 E mail info gamahealthcare com from the makers of Scan QR...