7.4 Transducer Cable Continuity Test
When difficulties are encountered with the ViscoSite transmitter and/or
transducer, a transducer cable continuity test can provide as much information
as to what may be the cause as it can to what is not the cause.
A continuity test of the wires should be carried out on the transducer cable using
a multimeter set to measure electrical resistance (
Ω
, Ohms.) Table 7.2 shows the
expected resistance (in Ohms) between specific pins of the terminal block with
and without the intrinsically safe barriers present.
Terminal Block Positions
Resistance (
Ω
)
No IS Barriers
(Ohms)
Resistance (
Ω
)
With IS Barriers
(Ohms)
J11
1 / 2
~70 to 100
~400 to 450
4 / 5
~70 to 100
~750 to 800
J10
1 / 2
~ 110
~160
1 / 3
~ 110
~160
2 / 3
<10
<50
4 / 5
~110
~160
4 / 6
~110
~160
5 / 6
<10
<50
Table 7.2: Transducer Cable Continuity Test
–
Expected Results
If the results of the continuity test do not match the expected results shown in
Table 7.2, contact the service department of Galvanic Applied Sciences Inc for
assistance. Refer to Section 7.5 for contact information.
7.5 Contact Galvanic Applied Sciences Inc.
In the event that the information in this section does not lead to a successful
diagnosis and resolution of an issue with the ViscoSite Viscometer as
implemented in your application, contact Galvanic Applied Sciences Inc.’s
Service Department.
The Service Department offers both phone/email technical support and on-site
technical service as required.
For Service and/or Assistance, contact:
Galvanic Applied Sciences Inc. USA
101 Billerica Ave
Building 5, Suite 104
North Billerica, MA 01862
USA
Phone: (978) 848-2701