The IP address and description can be selected via the drop down menus. The polling
analyzer data time can be set between 200 and 1000 ms. Press
Connect
when the
appropriate IP address and name are indicated.
Connect to Front Panel -
Allows for communication between the computer and the
chromatograph via the front Ethernet connection. This option is normally used when the
analyzer is not connected to a network.
Connect to Default Connection -
Automatically connects the computer to the system using
the IP indicated by the selected default check mark (Figure 4-12).
When the connection between the computer and the chromatograph is made, the
Select
Mode
dialog box (Figure 4-14) will be presented.
Figure 4-14: Select Mode Dialog Box
•
View
mode allows the user to collect data using the chromatograph but does not permit
changes to the analytical parameters.
•
Edit
allows the user to make changes to the analytical parameters of the analyzer. If Edit
mode is chosen, a dialog box will appear, prompting the user to enter a password. Edit
mode is described in Section 5.
•
Factory
mode is reserved for service engineers.
Summary of Contents for AccuChrome MA2963
Page 21: ...Revision 4 8 12 15 2018 Figure 2 7 The Intrinsically Safe Barrier ...
Page 24: ...Revision 4 11 12 15 2018 Figure 2 10 4 Minute Analysis Flow Diagram ...
Page 58: ...Revision 4 45 12 15 2018 Figure 4 24 Chart Commands ...
Page 65: ...Revision 4 52 12 15 2018 A sample report is shown in Figure 4 32 Figure 4 32 Sample Report ...
Page 80: ...Revision 4 67 12 15 2018 Figure 5 10 Streams Setup ...
Page 104: ...Revision 4 91 12 15 2018 Figure 8 2 Enron Mode Short Integers ...
Page 140: ...Revision 4 127 12 15 2018 ...
Page 141: ...Revision 4 128 12 15 2018 15 2 Valve Maintenance Instructions ...
Page 142: ...Revision 4 129 12 15 2018 ...