Revision 4
59
April 2020
spectrum is the same or very similar to the factory reference, this could indicate an
issue with plugging in the sample probe. Refer to Section 7.9 for details on how to clear
a plugged probe.
The
New Reference
button is used to make the analyzer use the currently displayed
transmission spectrum to calculate a new zero baseline without carrying out a full zero
calibration cycle. Highlight the New Reference button by using the F4 / F3 keys, then
pressing
↵
.
Pressing New Reference DOES NOT cause the analyzer to automatically go into
back purge mode. Be sure the analyzer is in back purge mode prior to using this
function otherwise the calculated analyzer results after calculating the new
reference WILL NOT BE VALID!
4.7
Absorbance Panel
The
Absorbance
panel shown in Figure 34 shows the currently calculated absorbance
spectrum as well as the current concentration of all of the analyzed components in the tail gas
stream.
Figure 34: Absorbance Panel
The
absorbance spectrum
is calculated by subtracting the current transmission spectrum from
the transmission spectrum that was obtained the last time the analyzer carried out a zero
calibration cycle OR when the New Reference button was pressed on the Spectrum panel,
whichever was most recent. The y-axis of this spectrum is absorbance units (AU). When the
analyzer is in back purge mode, the absorbance spectrum should be a horizontal line at an
absorbance of 0 AU. If it is not, and the analyzer is NOT in a fault condition, first confirm that
the current transmission spectrum in back purge is largely similar to the factory reference
spectrum, then immediately carry out a new zero calibration cycle OR use the New Reference
Summary of Contents for 943-TGX-CE
Page 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Page 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Page 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Page 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Page 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Page 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...