Revision 4
143
April 2020
Figure 89: Manual Override Page
In order to make any changes on this page,
Direct Peripheral Board Control
must first be
enabled by placing an X in the checkbox. Once this is enabled, the IO board will be under
manual control of the user and the
analyzer’s control computer will have no control of this board
until this direct board control is manually disabled.
The analyzer should be placed into back purge mode and set to Offline state prior to
performing any operations under Direct Board Control.
When work is complete on this page, the Direct Board Control Enable checkbox
must be unchecked manually to return the IO board to automatic control, otherwise
the analyzer will be unable to function normally.
12.3.1 Analog Output Calibration and Testing
Each
Analog Output
box allows the user to configure, calibrate, and test an analog output.
To configure/calibrate/test an Analog Output, follow the procedure below. This procedure is
identical for all four analog outputs:
1. Place a check mark in the
Calibrate Analog Output x
check box. The fields within the
box will become active.