i) When purging is complete, remove the oven compartment cover.
j) Switch valve V4 to the Off position. Ensure V5 is On. Switch V2 to the neutral position.
k) Remove the steam connection from the Probe Access ports and replace the sections of
tubing that connect the measurement cell to valves V1 and V3 to allow the back purge air
through the probe.
l) Replace the oven compartment cover. Ensure that the Fibre Optic Cables are still routed
in the indents to avoid damage when the oven compartment cover is secure.
m) Close and latch the oven cabinet door.
n) Allow the back purge to continue for five to ten minutes.
o) Perform a Manual Zero by clicking the Manual Zero button.
p) Remove the Zero mode by clicking the Manual Zero button again.
q) Click the On Line / Off Line to show ‘On Line’. The system will automatically restart itself.
The analyzer may show a fault and be in Manual (back purge) mode when
put On Line. This is due to a low oven temperature. Once the oven
temperature reaches its set point and the preset time delay expires, the
system will automatically restart.
7.10 Spectrometer Replacement
Consult Galvanic Applied Sciences for spectrometer replacement.
Summary of Contents for 943-TGS
Page 50: ...Revision 1 50 6 27 2017 Figure 4 14 Network Sub Panel ...
Page 94: ...Revision 1 94 6 27 2017 Figure 9 1 Mounting Detail ...
Page 95: ...Revision 1 95 6 27 2017 Figure 9 2 Oven Cabinet Detail ...
Page 96: ...Revision 1 96 6 27 2017 Figure 9 3 Control Cabinet Detail ...
Page 97: ...Revision 1 97 6 27 2017 Figure 9 4 Power Air Steam Requirements and Input Output Detail ...
Page 98: ...Revision 1 98 6 27 2017 Figure 9 5 Control Cabinet to Oven Cabinet Interconnect Detail ...
Page 99: ...Revision 1 99 6 27 2017 Figure 9 6 AC Wiring Detail ...
Page 100: ...Revision 1 100 6 27 2017 Figure 9 7 DC and Signal Wiring Detail ...
Page 101: ...Revision 1 101 6 27 2017 This page intentionally left blank ...